ETHICON INC. PRNVA BLU 36IN 3-0 D/A SH; SUTURE, NON ABSORBABLE, SYNTHETIC Back to Search Results
Model Number 3522H
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/14/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent a cardiovascular procedure on (b)(6) 2022 and suture was used.When opening the package, the surgeon tried to use then the suture became detached from the needle.It was not a control release needle.No adverse patient consequences were reported.Additional information was requested.
 
Additional Manufacturer Narrative
(b)(4).Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.-additional information was requested, and the following was obtained: please provide the lot number: rabbjr.The following information was requested, but unavailable: when was the needle detached/pulled off from the suture (during handling or during use on the patient)? please specify this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint #
=
> pc-(b)(4).Date sent to the fda: 08/11/2022 investigation summary: the product was returned for evaluation.Visual inspection was conducted on the returned sample.One opened sample product code 3522h was returned to ethicon for analysis.Upon visual analysis of the returned sample revealed that a clip off defect was observed in the sample.The barrel hole was examined under 20x magnification and revealed a remnant suture.In addition, the extreme suture was noted to be severely cut.As per returned conditions of the sample no functional test was performed.This defect is caused by excessive pressure on the suture during swaging of the needle and consequently breaking.Based on the information currently available, the clip-off was identified during the investigation of the sample received.This product issue will be addressed through ethicon quality system.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post-market surveillance.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRNVA BLU 36IN 3-0 D/A SH
Common Device Name
SUTURE, NON ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-SAN LORENZO PR
road 183, km. 8.3
san lorenzo 00754
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key14987576
Report Number2210968-2022-05372
Device Sequence Number16578828
Product Code GAW
UDI-Device Identifier10705031004450
UDI-Public10705031004450
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K001625
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2025
Device Model Number3522H
Device Catalogue Number3522H
Device Lot NumberRABBJR
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/14/2022
Supplement Date Received by Manufacturer07/28/2022
Initial Report FDA Received Date07/12/2022
Supplement Report FDA Received Date08/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/11/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-