| Model Number |
HST III SEAL (4.5MM), 5-PACK |
| Medical Device Problem Code |
Fitting Problem (2183)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/20/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Trackwise id: (b)(4).
|
| |
|
Event or Problem Description
|
|
Related to (b)(4).Summary: the hospital reported that during preparation for a coronary artery bypass procedure after an hs failure, hst iii seal (4.5mm), 5-pack the seal in the delivery sheath is kinked and cannot be loaded.The operation was completed with no problems using the new device and the patient is fine.
|
| |
|
Additional Manufacturer Narrative
|
|
Trackwise#: (b)(4).The lot # 25159573 history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.H3 other text : device not returned.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The device was returned to the factory for evaluation on 11/14/2022.An investigation was conducted on 12/14/2022.A visual inspection was conducted.Signs of clinical use and no evidence of blood was observed.The delivery device was returned inside the loading device with the white plunger not depressed and the blue safety lock on, which prevents the white plunger from being depressed.The delivery device was removed from the loading device with no physical or visual difficulties observed.The seal and tension spring assembly remained inside the loading device.The seal and tension spring assembly was removed from the loading device with no physical or visual difficulties observed.Measurements of the delivery device were taken; the inner diameter was measured at 0.197 inches, the outer diameter was measured at 0.221 inches (rm2036883).The length of the delivery tube was measured at 2.49 inches (mcv00004217).The measurement values recorded for the delivery tube were within the tolerance specifications.Based on the returned condition of the device as well as the investigation results and the reported failure "fitting problem" was confirmed.The lot # 25159573 history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.
|
| |
|