• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT GMBH ALINITY S ANTI-HCV REAGENT KIT; HEPATITIS C VIRUS ENCODED ANTIGENS (RECOMBINANT C100-3, HCR43)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT GMBH ALINITY S ANTI-HCV REAGENT KIT; HEPATITIS C VIRUS ENCODED ANTIGENS (RECOMBINANT C100-3, HCR43) Back to Search Results
Catalog Number 06P04-60
Medical Device Problem Code False Negative Result (1225)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/18/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.All available patient/donor information was included.Additional details are not available.
 
Event or Problem Description
A nonreactive alinity s anti-hcv result of 0.32 s/co was generated for an alinity s anti-hcv ii clinical trial donor re-entry sample (sid (b)(6).The donor was previously deferred during alinity s anti-hcv ii clinical trials.The nonreactive alinity s anti-hcv result of 0.32 s/co is considered to be false nonreactive.Donor re-entry process is conducted for samples that were discrepant during initial clinical trial testing for the alinity s anti-hcv ii assay.Re-entry testing: (b)(6) 2022.Sid (b)(6).Alinity s anti-hcv result: nonreactive 0.32 s/co.Alinity s anti-hcv ii result: repeat reactive 13.59, 13.44, 13.30 s/co.Nat result: nonreactive.Ortho result: nonreactive.Initial testing: (b)(6) 2021.Sid (b)(6).Alinity s anti-hcv result: nonreactive 0.26 s/co.Alinity s anti-hcv ii result: repeat reactive 11.53, 12.06, 12.02 s/co.Nat result: negative.Roche anti-hcv ii result: reactive.No impact to donor or patient management was reported.
 
Additional Manufacturer Narrative
Trending review determined no adverse trend for the issue for the product.Historical complaint review determined there is normal complaint activity for the lot number.Device history record review did not show any potential non-conformances, or deviations.Labeling was reviewed and found to adequately address the issue under review.To evaluate clinical sensitivity, 112 replicates of two sensitivity panels (anti-hcv panel a (core only) and anti-hcv panel b (ns3 only)) were tested.The sensitivity panels met specifications.No false non-reactive results were obtained.Additionally, two commercially available seroconversion panels (zeptometrix hcv seroconversion panels hcv9045 and hcv9047) were tested.Alinity s anti-hcv reagent, ln 6p04-60, lot 35225be00 detected the same first bleed as reactive compared to historical results for alinity s anti-hcv.Further, additional replicates of the positive control were tested which all met specifications.Based on the investigation, no deficiency for lot number 35225be00 was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALINITY S ANTI-HCV REAGENT KIT
Common Device Name
HEPATITIS C VIRUS ENCODED ANTIGENS (RECOMBINANT C100-3, HCR43)
Manufacturer (Section D)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM  65205
Manufacturer (Section G)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM   65205
Manufacturer Contact
siobhan wright
lisnamuck
post market surveillance
longford N39 E-932
EI   N39 E932
433331157
MDR Report Key14987799
Report Number3002809144-2022-00241
Device Sequence Number1674550
Product Code QHM
UDI-Device Identifier00380740117191
UDI-Public00380740117191
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
BL 125677
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/04/2022
Device Catalogue Number06P04-60
Device Lot Number35225BE00
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/27/2022
Supplement Date Received by Manufacturer08/01/2022
Initial Report FDA Received Date07/12/2022
Supplement Report FDA Received Date08/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/03/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ALNTY S SYSTEM, 06P16-01, (B)(6); ALNTY S SYSTEM, 06P16-01, (B)(6)
Patient SexUnknown
-
-