SAKURA FINETEK USA, INC. TISSUE-TEK AUTOSECTION(R) AUTOMATED MICROTOME Back to Search Results
Model Number 5010
Medical Device Problem Code Device Handling Problem (3265)
Health Effect - Clinical Code Laceration(s) (1946)
Date of Event 06/21/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This is a user error by sakura employee.Employee did not follow the instruction as per operating manual and did not ensure the blade guard was in place before removing the cassette.
 
Event or Problem Description
Sakura finetek was notified of an incident on (b)(6) 2022 which occurred on (b)(6) 2022 with the autosection instrument, serial no.(b)(4).Incident occurred during installation of instrument.While setting up the autosection instrument, sakura field application specialist noticed the blocks that the client used were bigger than the normal cassettes used; therefore, sakura employee put one of the client's cassette in the unit to ensure it would clear the blade holder.Employee pressed the align button and realized he did not want to cut into the client's block, so the employee tried to grab the cassette before the unit started.The unit started and brought employee's finger on the blade causing a deep cut across the tip of employee's left thumb.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TISSUE-TEK AUTOSECTION(R) AUTOMATED MICROTOME
Common Device Name
AUTOSECTION
Manufacturer (Section D)
SAKURA FINETEK USA, INC.
1750 west 214th street
torrance CA 90501
Manufacturer (Section G)
SAKURA FINETEK USA, INC.
1750 west 214th st
torrance CA 90501
Manufacturer Contact
solmaz shaida
1750 west 214th street
torrance, CA 90501
3109727800
MDR Report Key14988015
Report Number2083544-2022-00001
Device Sequence Number1535080
Product Code IDO
UDI-Device Identifier00615233079350
UDI-Public00615233079350
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Remedial Action Other
Type of Report Initial
Report Date (Section B) 06/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Other
Device Model Number5010
Device Catalogue Number5010
Was Device Available for Evaluation? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/21/2022
Initial Report FDA Received Date07/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/01/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient SexUnknown
-
-