• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® X-FORCE® BALLOON DILATION CATHETER; X-FORCE URETERAL DILATION BALLOON CATHETER KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® X-FORCE® BALLOON DILATION CATHETER; X-FORCE URETERAL DILATION BALLOON CATHETER KIT Back to Search Results
Model Number 998706
Medical Device Problem Codes Burst Container or Vessel (1074); Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/22/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during inspection of the balloon dilation catheter prior to operation, the doctor found that the balloon leaked water when being inflated using a pump.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during inspection of the balloon dilation catheter prior to operation, the doctor found that the balloon leaked water when being inflated using a pump.Per additional information via email from ibc on 27sep2022, it was confirmed that leak was coming from the balloon.
 
Additional Manufacturer Narrative
The reported event is confirmed, cause unknown.The used x-force balloon dilation catheter was returned in the opened original packaging.An evaluation was completed by futurematrix on 05dec2022.Potential root cause for this event could be, "shaft wall too thin".No manufacturing issues or non-conformances were noted during review of the dhr that could have caused or contributed to the reported event.Based on the results of the investigation no additional action is required at this time.The instructions for use were found adequate and state the following: "do not exceed the recommended rated burst pressure (rbp) for this device." "only a physician who has an understanding of the clinical applications, technical principles and associated risks of balloon dilation within the urinary tract should use this device." "the x-force® balloon dilation catheter is a sterile, single use device.Carefully inspect the catheter and the sterile packaging for signs of damage that may have occurred during shipment.Do not use the product if damage is evident." "store the balloon dilation catheter in a cool, dry, dark place.Do not store near radiation or ultraviolet light sources as these may damage product materials." h11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® X-FORCE® BALLOON DILATION CATHETER
Common Device Name
X-FORCE URETERAL DILATION BALLOON CATHETER KIT
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
juan velez
8195 industrial blvd
covington 30014
7707846100
MDR Report Key14988437
Report Number1018233-2022-05353
Device Sequence Number19303001
Product Code EZN
UDI-Device Identifier10801741052740
UDI-Public(01)10801741052740
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K050875
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/29/2022
Device Model Number998706
Device Catalogue Number998706
Device Lot NumberBMEPFM18
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/22/2022
Supplement Date Received by Manufacturer12/19/2022
Initial Report FDA Received Date07/12/2022
Supplement Report FDA Received Date12/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/19/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-