• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION OPTICROSS 35; CATHETER, ULTRASOUND, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION OPTICROSS 35; CATHETER, ULTRASOUND, INTRAVASCULAR Back to Search Results
Model Number H7493932800350
Medical Device Problem Code Material Perforation (2205)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/20/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product code: obj, itx.
 
Event or Problem Description
It was reported that device perforation occurred.The patient presented for a lower extremity venous intervention.When the opticross35 imaging catheter was being loaded onto the guidewire, the wire punctured through the lumen of the catheter and exited close to the distal end of the catheter and not through the wire lumen hub on the proximal end.The procedure was completed with another of the same device.There were no patient complications.
 
Additional Manufacturer Narrative
The device was returned for analysis.Visual inspection revealed a kink in the sheath assembly.During microscopic inspection noted pitting and degradation of the transducer housing consistent with saline damage which occurs post-use of the device and is not related to the original device failure.A perforation was observed in the sheath near the kinked section.
 
Event or Problem Description
It was reported that device perforation occurred.The patient presented for a lower extremity venous intervention.When the opticross35 imaging catheter was being loaded onto the guidewire, the wire punctured through the lumen of the catheter and exited close to the distal end of the catheter and not through the wire lumen hub on the proximal end.The procedure was completed with another of the same device.There were no patient complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTICROSS 35
Common Device Name
CATHETER, ULTRASOUND, INTRAVASCULAR
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key14988671
Report Number2134265-2022-07824
Device Sequence Number19251099
Product Code OBJ
UDI-Device Identifier08714729984542
UDI-Public08714729984542
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K200733
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/11/2022
Device Model NumberH7493932800350
Device Catalogue NumberH7493932800350
Device Lot Number0028876115
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/20/2022
Supplement Date Received by Manufacturer08/10/2022
Initial Report FDA Received Date07/12/2022
Supplement Report FDA Received Date08/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/11/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-