Model Number N/A
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/16/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported during a shoulder orthoscopy surgery, when the surgeon tried to insert the anchor in desired hole, the anchor broke.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Additional Manufacturer Narrative
(b)(4).Report source: foreign: india.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected d10 : quantity 3 part # 912031 jgrknt 1.5mm #2 mb sngl 2 mb sngl lot # 0002392612 visual examination of the returned product identified that the anchor is fractured where the sutures are inserted and the base remains at the tip of the inserter.The fractured piece of the anchor was not returned for evaluation.The remaining piece of anchor is in place at the tip of the inserter with the sutures running over the top of it.The sutures remain threaded through the shaft, down to the handle with the suture card still in place inside the handle.Medical records were not provided.Device history records were reviewed and no discrepancies related to the reported event were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUATTRO GL ANCHOR 2.9MM
Common Device Name
FASTENER, FIXATION
Manufacturer (Section D)
CAYENNE MEDICAL
16597 n 92nd street
scottsdale AZ 85260
Manufacturer (Section G)
CAYENNE MEDICAL
16597 n 92nd street
scottsdale AZ 85260
Manufacturer Contact
jennifer rapsavage
56 e. bell dr
warsaw, IN 46582
5745260384
MDR Report Key14989321
Report Number3006108336-2022-00022
Device Sequence Number13923529
Product Code MBI
UDI-Device Identifier00887868270352
UDI-Public(01)00887868270352(17)221024(10)64519-1
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K131325
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/24/2022
Device Model NumberN/A
Device Catalogue NumberCM-9329
Device Lot Number64519-1
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/20/2022
Supplement Date Received by Manufacturer08/30/2022
Initial Report FDA Received Date07/12/2022
Supplement Report FDA Received Date09/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/24/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10.
Patient Age23 YR
Patient SexMale
-
-