• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INTELLANAV MIFI OPEN-IRRIGATED; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INTELLANAV MIFI OPEN-IRRIGATED; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER Back to Search Results
Model Number 87047
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/31/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was received at boston scientific for analysis.Visual inspection using an electron microscope revealed a foreign material with a yellow-brown color inside the tip of the device.It was not possible to identify the nature of the organic residue.In the flow, the device presented a partial obstruction in some of the irrigation holes.The unit passed functional, leak, and continuity tests.The device did not pass the electrical test since there was evidence of noise.Analysis of the returned product revealed that the unit showed no evidence or defects that could have contributed to the tracking/accuracy issue event, consequently, not confirming this reported failure.Based on the information provided, there is no evidence that the device was used in a manner inconsistent with the labelled indications.
 
Event or Problem Description
Reportable based on analysis completed on 15 jun 2022.During the ischemic ventricular tachycardia ablation, an intellanav mifi oi catheter was selected for use.It was reported that there was noise and static on all surfaces and intracardiac electrograms.This issue occurred while manipulating.The catheter had intermittent visibility on the rhythmia.The mifi oi catheter was replaced, and the case was successfully completed.No patient complications were reported.Product analysis revealed there was a foreign material in the distal tip and in the flow.The device presented a partial obstruction in some of the irrigation holes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLANAV MIFI OPEN-IRRIGATED
Common Device Name
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
dc a330
saint paul, MN 55112
6515826168
MDR Report Key14989546
Report Number2134265-2022-07848
Device Sequence Number19266836
Product Code OAD
UDI-Device Identifier08714729938378
UDI-Public08714729938378
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P150005/S017
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/07/2024
Device Model Number87047
Device Catalogue Number87047
Device Lot Number0028819477
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/18/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/15/2022
Initial Report FDA Received Date07/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/07/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-