• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR HI-TORQUE PILOT GUIDE WIRE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR HI-TORQUE PILOT GUIDE WIRE Back to Search Results
Model Number 1010480-H
Medical Device Problem Codes Difficult to Remove (1528); Deformation Due to Compressive Stress (2889); Difficult to Advance (2920)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/15/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history identified no other incidents from this lot.The investigation determined the reported difficulties appear to be related to the operational context of the procedure.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
 
Event or Problem Description
It was reported that the procedure was to treat a 90% stenosed lesion in the left circumflex (cx) artery with moderate calcification and moderate tortuosity.The 014 hi-torque pilot guide wire (gw) was advanced beyond the target lesion with resistance noted from the lesion.During pullback, the proximal shaft of the wire kinked.The gw was removed from the patient with resistance again from the lesion.There was not adverse patient effect and no clinically significant delay reported.The procedure was continued with another pilot gw.No additional information has been provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HI-TORQUE PILOT GUIDE WIRE
Common Device Name
GUIDE WIRE
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005737652
road no.2 km 58.0 cruce dávila
barceloneta PR 00617
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key14990070
Report Number2024168-2022-07559
Device Sequence Number1674605
Product Code DQX
UDI-Device Identifier08717648073854
UDI-Public08717648073854
Combination Product (Y/N)N
Initial Reporter CountryTH
PMA/510(K) Number
K101116
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2022
Device Model Number1010480-H
Device Catalogue Number1010480-H
Device Lot Number0111071
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/20/2022
Initial Report FDA Received Date07/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/10/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-