• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR GRAFTMASTER RX CORONARY STENT GRAFT SYSTEM; CORONARY STENT DELIVERY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR GRAFTMASTER RX CORONARY STENT GRAFT SYSTEM; CORONARY STENT DELIVERY SYSTEM Back to Search Results
Model Number 1012581-19
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Code Perforation of Vessels (2135)
Date of Event 06/10/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the procedure was to treat the distal left main coronary artery.Lithotripsy was performed and a xience skypoint stent and a non-abbott stent were implanted.A significant perforation was noted after the non-abbott stent placement.A 4.0x16 mm graftmaster covered stent was implanted and partially sealed the perforation after the patient's blood pressure dropped.A 3.5x19 mm graftmaster covered stent was implanted and partially sealed the perforation after cpr was started.When the vessel was initially stented, it was reported that the vessel "crumbled" and could not sustain the covered stent.The patient expired; however, it was reported the graftmaster covered stents did not cause or contribute to complications or adverse events as they were used purely as rescue devices.No additional information was provided.
 
Additional Manufacturer Narrative
The stent remains in patient.The device will not be returned for evaluation.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.The additional graftmaster device is being filed under a separate medwatch report number.
 
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents.The reported patient effect of perforation, as listed in the graftmaster rx coronary stent graft system, instructions for use (ifu), is a known patient effect that may be associated with use of a coronary stent in native coronary arteries.The investigation was unable to determine a conclusive cause for the reported failure to seal.A conclusive cause for the reported patient effects, and the relationship to the product, if any, cannot be determined; however, the subsequent treatment appears to be related to the operational context of the procedure.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
 
Event or Problem Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GRAFTMASTER RX CORONARY STENT GRAFT SYSTEM
Common Device Name
CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key14990792
Report Number2024168-2022-07568
Device Sequence Number19223155
Product Code MAF
UDI-Device Identifier08717648176401
UDI-Public08717648176401
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
H000001
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2023
Device Model Number1012581-19
Device Catalogue Number1012581-19
Device Lot Number1042941
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/22/2022
Supplement Date Received by Manufacturer07/22/2022
Initial Report FDA Received Date07/12/2022
Supplement Report FDA Received Date08/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/29/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GRAFTMASTER US RX 4.00X16
Outcome Attributed to Adverse Event Required Intervention;
Patient Age84 YR
Patient SexMale
Patient Weight73 KG
-
-