ETHICON INC. PRLNE BLU 18IN 7-0 D/A BV175-6 EP; SUTURE, NONABSORBABLE, SYNTHETIC Back to Search Results
Model Number EP8734H
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/13/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent a cabg procedure (b)(6) 2022 and suture was used.During the procedure, the inner needle at both ends of the product detached the moment the surgeon held it.A new package was used for vascular anastomosis, and after continuous suturing on lad, when the surgeon was about to pass the needle-suture to a scrub nurse, the needle detached from the suture again and fell into the patient's pleural cavity.It took several minutes to find the fallen needle and there was no residue in the patient's body.These were not control release needles.No adverse patient consequences were reported.Additional information was requested.
 
Additional Manufacturer Narrative
(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The following information was requested, but unavailable: was additional dissection required in other tissues other than the target tissue to remove the needle? if yes, please describe: no further information is available.Was the patient treatment or post-operative care altered in anyway due to the prolonged surgery time? if yes, please explain: no further information is available.Was there any patient impact (ie.Additional treatment required) and/or adverse patient outcome due to the prolonged surgery? no further information is available.Was the procedure completed successfully? no further information is available.What is the patient¿s current status? no further information is available.No further information will be provided." additional information was requested, and the following was obtained via: no patient damage occurred with the first needle used.No surgical action was taken during the search for the second needle (this was the needle-suture which was used for vascular anastomosis).The patient has been discharged from the hospital already and has no complaints about the event.Events reported via: 2210968-2022-05394.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 8/9/2022.Additional information: d9, h3, h4, h6.H6 component code: g07002 no device problem found a manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Additional information was requested, but unavailable: was additional dissection required in other tissues other than the target tissue to remove the needle? if yes, please describe no further information is available.Was the patient treatment or post-operative care altered in anyway due to the prolonged surgery time? if yes, please explain.No further information is available.Was there any patient impact (ie.Additional treatment required) and/or adverse patient outcome due to the prolonged surgery? no further information is available.Was the procedure completed successfully? no further information is available.What is the patient¿s current status? no further information is available.No further information will be provided." additional information was received: no patient damage occurred with the first needle used.No surgical action was taken during the search for the second needle (this was the needle-suture which was used for vascular anastomosis).The patient has been discharged from the hospital already and has no complaints about the event.H3 investigational summary: the product was returned to ethicon for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample revealed that one suture piece of product code ep8734h was received for analysis.In order to evaluate the conditions of the returned sample, the needle was not received for evaluation, and during the visual inspection of the suture, the ends were observed cut.After further evaluation crimping marks were observed on the surface suture possibly from a surgical instrument.As with any device, care should be taken to avoid damage to the strand when handling.Avoid the crushing or crimping action of the surgical instruments, such as needle holders and forceps.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.The condition of the sample received indicates improper handling of the device.It should be noted that as part of our quality process, each batch is randomly inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Additional h3 investigational summary: the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that two opened samples of product code ep8734h were received for analysis.In order to evaluate the conditions of the returned sample, the swage and attachment area was noted to be as expected.The sutures were dispensed without problems and examined along the strand to detect any issues related no defects were observed during evaluation.Functional test was performed, and the pull force result were above the minimum requirements.The event described could not be confirmed as the device performed without any defect noted.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.It should be noted that as part of our quality process, each batch is randomly inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Additional h3 investigational summary: the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that two opened boxes with thirty-six and nine packets ea.Of product code ep8734h were returned with the packaging closed.Upon initial inspection, of the packets, no external damages were observed.As per the sampling plan, a visual inspection was performed on thirteen samples, the packets were opened, and no defects were detected.The swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along the strand to detect any issues related no defects were observed during the evaluation.Functional test was performed, and the pull force result were above the minimum requirements.The event described could not be confirmed as the device performed without any defect noted.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.It should be noted that as part of our quality process, each batch is randomly inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Events reported via: 2210968-2022-05395 and 2210968-2022-05394.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRLNE BLU 18IN 7-0 D/A BV175-6 EP
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-SAN LORENZO PR
road 183, km. 8.3
san lorenzo PR
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key14991029
Report Number2210968-2022-05395
Device Sequence Number1613906
Product Code GAW
UDI-Device Identifier10705031206663
UDI-Public10705031206663
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K133356
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEP8734H
Device Catalogue NumberEP8734H
Device Lot NumberRMBACR
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/16/2022
Supplement Date Received by Manufacturer07/28/2022
Initial Report FDA Received Date07/12/2022
Supplement Report FDA Received Date08/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-