| Model Number |
36703 |
| Medical Device Problem Code |
Therapeutic or Diagnostic Output Failure (3023)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/19/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that insufficient roll off occurred.A soloist single needle electrode was selected for the radiofrequency ablation of an osteoid osteoma.A 13 gauge x 10cm bone biopsy needle was placed on the nidus and the soloist needle was passed through it.Ablation was started and ran for 15 minutes.After 15 minutes, impedance was not achieved and ablation was restarted.Impedance was unable to be achieved during the second attempt.The soloist needle was repositioned but impedance could not be achieved.The needle was removed and the procedure was completed with another device.There were no patient complications reported.
|
| |
|
Event or Problem Description
|
|
It was reported that insufficient roll off occurred.A soloist single needle electrode was selected for the radiofrequency ablation of an osteoid osteoma.A 13 gauge x 10cm bone biopsy needle was placed on the nidus and the soloist needle was passed through it.Ablation was started and ran for 15 minutes.After 15 minutes, impedance was not achieved and ablation was restarted.Impedance was unable to be achieved during the second attempt.The soloist needle was repositioned but impedance could not be achieved.The needle was removed and the procedure was completed with another device.There were no patient complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: one electrode needle was returned for analysis and its respective connection cable.No visual damages defects were observed on the electrode needle.An electrical evaluation was performed by measuring the continuity, the electrode was able to conduct current indicating proper connectivity.
|
| |
|
Search Alerts/Recalls
|