• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SOLOIST; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SOLOIST; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number 36703
Medical Device Problem Code Therapeutic or Diagnostic Output Failure (3023)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/19/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that insufficient roll off occurred.A soloist single needle electrode was selected for the radiofrequency ablation of an osteoid osteoma.A 13 gauge x 10cm bone biopsy needle was placed on the nidus and the soloist needle was passed through it.Ablation was started and ran for 15 minutes.After 15 minutes, impedance was not achieved and ablation was restarted.Impedance was unable to be achieved during the second attempt.The soloist needle was repositioned but impedance could not be achieved.The needle was removed and the procedure was completed with another device.There were no patient complications reported.
 
Event or Problem Description
It was reported that insufficient roll off occurred.A soloist single needle electrode was selected for the radiofrequency ablation of an osteoid osteoma.A 13 gauge x 10cm bone biopsy needle was placed on the nidus and the soloist needle was passed through it.Ablation was started and ran for 15 minutes.After 15 minutes, impedance was not achieved and ablation was restarted.Impedance was unable to be achieved during the second attempt.The soloist needle was repositioned but impedance could not be achieved.The needle was removed and the procedure was completed with another device.There were no patient complications reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr: one electrode needle was returned for analysis and its respective connection cable.No visual damages defects were observed on the electrode needle.An electrical evaluation was performed by measuring the continuity, the electrode was able to conduct current indicating proper connectivity.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOLOIST
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key14991436
Report Number2134265-2022-07756
Device Sequence Number1597090
Product Code GEI
UDI-Device Identifier08714729122487
UDI-Public08714729122487
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K053128
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/25/2025
Device Model Number36703
Device Catalogue Number36703
Device Lot Number0028768923
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/19/2022
Supplement Date Received by Manufacturer10/26/2022
Initial Report FDA Received Date07/12/2022
Supplement Report FDA Received Date11/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/26/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age28 YR
Patient SexFemale
Patient Weight60 KG
-
-