• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CHRISTOPH MIETHKE GMBH & CO KG M.BLUE 5 SYS W/PED.CONTROL RESERVOIR; HYDROCEPHALUS MANAGEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CHRISTOPH MIETHKE GMBH & CO KG M.BLUE 5 SYS W/PED.CONTROL RESERVOIR; HYDROCEPHALUS MANAGEMENT Back to Search Results
Model Number FX816T
Medical Device Problem Codes Mechanical Problem (1384); Infusion or Flow Problem (2964)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 05/24/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a m.Blue (part # fx816t) was implanted during a procedure performed on an unknown date.According to the complainant, the shunt system was believed to be operated in overdrainage and had adjustment difficulties.The patient underwent a revision procedure performed on (b)(6) 2022.The complaint device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.
 
Additional Manufacturer Narrative
Investigation: visual inspection: scratches on the outer housing of the valve were observed through the visual inspection.No significant deformations or damage were detected.Permeability test: a permeability test has shown that the valve is permeable.Adjustment test: the m.Blue valve was tested and is not adjustable throughout the normal range.Braking force and brake function test: the brake functionality test has shown that the brake function is operational, however the braking force cannot be measured due to the non-adjustability of the valve.(see section above).Computer controlled test: to investigate the claim of over-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The results show that the valve is operating within the accepted tolerance in the horizontal but not in the vertical position.Internal inspection: after dismantling of the valve, deposits were found in m.Blue.Results: for the sake of thoroughness and transparency, we would like to point out that a permeability test and adjustability test was already carried out after the revision at the hospital.Based on our investigation, we confirm that the valve shows an increased flow of liquid and a non-adjustability, likely due to the deposits observed inside the valve.As described in scientific literature, the problem encountered is one of the known, inevitable risks of hc-therapy by shunt implants.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M.BLUE 5 SYS W/PED.CONTROL RESERVOIR
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key14992877
Report Number3004721439-2022-00191
Device Sequence Number1880897
Product Code JXG
UDI-Device Identifier04041906504526
UDI-Public4041906504526
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K192266
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/25/2024
Device Model NumberFX816T
Device Catalogue NumberFX816T
Device Lot Number20042725
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/13/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/13/2022
Initial Report FDA Received Date07/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/05/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age4 YR
Patient SexMale
-
-