| Model Number |
29MJ-501 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
06/15/2022
|
|
Type of Reportable Event
|
Death
|
|
Additional Manufacturer Narrative
|
|
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.The date of death is estimated as on (b)(6) 2022, based on the reported date of event.
|
| |
|
Event or Problem Description
|
|
It was reported that on 29mm mitral masters valve was selected for implant on (b)(6) 2022.During the procedure a leaflet of the masters valve broke.No further information was provided about the procedure.The patient's status was noted as deceased.The date and cause of death is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
The previous report was submitted on 9/22/2022 and failed ack 3.This report is being resubmitted at the request of the cesub help desk due to an infrastructure issue at fda.An event of the valve breaking during rotation of the valve and patient death was reported.The investigation confirmed that the valve leaflet and orifice were fractured.Three images were received from the field, which appeared to show a valve with one leaflet dislodged and fractured, consistent with the received device.Information from the field reported that the valve breaking was unrelated to the patient's evolution.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.While the root cause of the leaflet fracture could not be conclusively determined, there was no evidence of material defect in the carbon coating that may have caused or contributed to the dislodged leaflet.There is no indication of a product quality issue with regards to manufacture, design, or labeling.
|
| |
|
Event or Problem Description
|
|
Subsequent to the previously filed report, additional information was received that the valve broke during valve implantation when an attempt was made to rotate.The valve and all pieces were removed from the patient a new unknown valve was used.
|
| |
|