The catalog number identified in section d4 has not been cleared in the us but is similar to the x-port isp implantable port, groshong single-lumen, 8f products that are cleared in the us.The pro code and 510 k number for the x-port isp implantable port, groshong single-lumen, 8f products are identified in d2 and g4.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.(expiration date: 01/2024).
|
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the x-port isp implantable port, groshong single-lumen, 8f products that are cleared in the us.The pro code and 510 k number for the x-port isp implantable port, groshong single-lumen, 8f products are identified in d2 and g4.H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one x-port mri isp implantable port attached to a groshong catheter was returned for evaluation.Functional, gross visual, tactile, and microscopic visual evaluations were performed.The investigation is inconclusive for the reported fluid leak issue, as the exact circumstances at the time of the reported event cannot be verified, and the reported event could not be reproduced in the lab.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: (expiration date: 01/2024).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|