• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 X-PORT ISP IMPLANTABLE PORT, GROSHONG SINGLE-LUMEN, 8F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 X-PORT ISP IMPLANTABLE PORT, GROSHONG SINGLE-LUMEN, 8F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR Back to Search Results
Catalog Number 7707540J
Device Problem Fluid/Blood Leak (1250)
Patient Problem Extravasation (1842)
Event Date 06/16/2022
Event Type  Injury  
Event Description
It was reported that approximately twenty four days post port placement procedure, the port allegedly had subcutaneous leak.It was further reported that the port was removed.Reportedly patient experienced extravasation.The current status of the patient is unknown.
 
Manufacturer Narrative
The catalog number identified in section d4 has not been cleared in the us but is similar to the x-port isp implantable port, groshong single-lumen, 8f products that are cleared in the us.The pro code and 510 k number for the x-port isp implantable port, groshong single-lumen, 8f products are identified in d2 and g4.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.(expiration date: 01/2024).
 
Event Description
It was reported that twenty four days post port placement procedure, the port allegedly had subcutaneous leak.It was further reported that the port was removed.Reportedly patient experienced extravasation.The current status of the patient is unknown.
 
Manufacturer Narrative
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the x-port isp implantable port, groshong single-lumen, 8f products that are cleared in the us.The pro code and 510 k number for the x-port isp implantable port, groshong single-lumen, 8f products are identified in d2 and g4.H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one x-port mri isp implantable port attached to a groshong catheter was returned for evaluation.Functional, gross visual, tactile, and microscopic visual evaluations were performed.The investigation is inconclusive for the reported fluid leak issue, as the exact circumstances at the time of the reported event cannot be verified, and the reported event could not be reproduced in the lab.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: (expiration date: 01/2024).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X-PORT ISP IMPLANTABLE PORT, GROSHONG SINGLE-LUMEN, 8F
Type of Device
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key14997765
MDR Text Key295768709
Report Number3006260740-2022-02663
Device Sequence Number1
Product Code LJT
UDI-Device Identifier00801741098246
UDI-Public(01)00801741098246
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K022983
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/13/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number7707540J
Device Lot NumberREFV0404
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/05/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/05/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/31/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-