Catalog Number 816047100
Medical Device Problem Code False Negative Result (1225)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/16/2022
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported weak positive to false negative reactions with biotestcell-a2 when they were expected to be clearly positive.The customer stated that she observed this issue when performing manual qc with immucor core qc kit.The customer did not provide the complaint sample biotestcell-a2 for investigational testing.Therefore; our quality control laboratory tested their retention sample of the supposedly defective lot with different donor samples for potency and specificity.All acceptance criteria were met.All positive and negative reactions were correct.We did not observe any false negative respectively weakened reaction.A review of the batch record documentation showed no irregularities, which might have negatively affected the quality of the allegedly defective lot.The complaint investigation showed that the customer did not use the correct ifu and did not realize the statement under the test procedure to incubate for 15 to 30 minutes if necessary.The customer incubated for an additional 15 minutes and the reaction came up as 2+ positive.The customer informed us that she received the correct results after following the instruction for use and incubating the test for 15 minutes.Based on the investigation and the information of the customer the complaint was classified as unjustifie.
 
Additional Manufacturer Narrative
This is our combined intital and final report on this incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REAGENT RED BLOOD CELLS BIOTESTCELL A2
Common Device Name
BIOTESTCELL A2
Manufacturer (Section D)
BIO-RAD MEDICAL DIAGNOSTICS GMBH
industriestrasse 1
dreieich, hessen 63303
GM  63303
Manufacturer (Section G)
BIO-RAD MEDICAL DIAGNOSTICS GMBH
industriestrasse 1
dreieich, hessen 63303
GM   63303
Manufacturer Contact
martina benkert
industriestrasse 1
dreieich, hessen 63303
GM   63303
MDR Report Key14998511
Report Number9610824-2022-00036
Device Sequence Number11233020
Product Code QHT
UDI-Device Identifier07611969952625
UDI-Public(01)07611969952625(17)220801(10)9222011-00
Combination Product (Y/N)N
Initial Reporter StateID
Initial Reporter CountryUS
PMA/510(K) Number
125207
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/01/2022
Device Catalogue Number816047100
Device Lot Number9222011-00
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/16/2022
Initial Report FDA Received Date07/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BIOTESTCELL A2, LOT 9218011-00; IMMUCOR QC A2, LOT 15122