• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 3.5 VA LCKNG SCR SLF-TPNG/SD/80; PLATE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 3.5 VA LCKNG SCR SLF-TPNG/SD/80; PLATE, FIXATION, BONE Back to Search Results
Model Number 02.127.180
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Nonunion/Delayed-union Bone Fracture (2369)
Date of Event 06/13/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported on (b)(6) 2022, that there was removal of hardware and revision.Non union right femur.All hardware was removed without complication.There were no broken implants.Procedure was completed successfully.Patient was doing well after post op.This report is for one (1) 3.5 va lckng scr slf-tpng/sd/80.This is report 2 of 4 for complaint (b)(4).
 
Additional Manufacturer Narrative
Additional narrative: additional product code: hwc.Complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3.5 VA LCKNG SCR SLF-TPNG/SD/80
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
WERK GRENCHEN (CH)
solothurnstrasse 186
grenchen
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key14998922
Report Number2939274-2022-02609
Device Sequence Number16562635
Product Code HRS
UDI-Device Identifier10886982043382
UDI-Public(01)10886982043382
Combination Product (Y/N)N
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
K121601
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number02.127.180
Device Catalogue Number02.127.180
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/14/2022
Initial Report FDA Received Date07/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age45 YR
Patient SexFemale
Patient Weight70 KG
-
-