DEPUY SPINE INC EXPEDIUM SPINE SYSTEM ERGONOMIC T-HANDLE; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 279702150
Device Problems Device-Device Incompatibility (2919); Appropriate Term/Code Not Available (3191)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/14/2022
Event Type  malfunction  
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Complainant part is not expected to be returned for manufacturer review/investigation.Initial reporter is a synthes employee.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from australia reports an event as follows: it was reported that after a procedure on (b)(6) 2022, the head adjuster attached to the t-handle was extremely difficult to separate.There was no patient consequence.This report is for a expedium spine system ergonomic t-handle.This is report 1 of 1 for (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: investigation summary : an analysis of the product could not be performed since a physical sample was not received for evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.However, if the product is received at a later date, the investigation will be updated as applicable.Device history lot: part #279702150vsf & lot # 0405gs combination is not valid, and device is not being returned, therefore, dhr could not be completed.If the device is returned or lot number can be confirmed,the dhr will be revisited.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPEDIUM SPINE SYSTEM ERGONOMIC T-HANDLE
Type of Device
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
DEPUY SPINE INC
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
chemin-blanc 38
le locle 02400
SZ   02400
3035526892
MDR Report Key14999237
MDR Text Key296248246
Report Number1526439-2022-01022
Device Sequence Number1
Product Code LXH
UDI-Device Identifier10705034198538
UDI-Public(01)10705034198538
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/13/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number279702150
Device Catalogue Number279702150
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/15/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-