Model Number 101041361
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 06/15/2022
Type of Reportable Event Serious Injury
Event or Problem Description
On (b)(6) 2022, the customer alleged to medela llc while using her pump in style maxflow her pump stopped suctioning.Additionally the customer alleged she developed mastitis and was prescribed antibiotics.
 
Additional Manufacturer Narrative
Customer service conducted troubleshooting with the customer, including verifying customer was using correct kit components, verifying customer was washing parts according to instruction for use manual, verifying customer was using dry parts while pumping and verifying customer was not washing tubing.When customer was asked if she had milk backup with her pump she indicated yes.A replacement pump, tubing and connectors were sent to the customer, and the return of her original pump was requested back for testing.In follow up with a complaint handler on (b)(6) 2022, the customer indicated that she developed mastitis three weeks after she began using her pump.She was prescribed an antibiotic, and she indicated that her mastitis was healed.The device was not returned to medela, only the customer's tubing was returned; therefore, it was evaluated with a medela lab pump on (b)(6) 2022.The tubing passed suction and cycle specifications while using the medela lab pump and kit.Also identified in the evaluation was human milk residue in the tubing.The customer's report of low suction could not be confirmed.Based on the results of our internal investigation (reference number (b)(4)), it cannot be definitively concluded that the pump caused or contributed to the customer¿s mastitis.The estimated incidence of mastitis in lactating women, whether using a breast pump or not, according to published clinical literature can be as high as 33%.In fact, clinical guidelines suggest the use of a breast pump to facilitate withdrawal of breast milk during bouts of mastitis.The complaint rate of mastitis across all reported failures, across all medela breast pumps, is (b)(4) for the period of (b)(6) 2013 to (b)(6) 2017.Mastitis is usually a benign, self-limiting infection with few consequences for the suckling infant.The risk of mastitis is higher among women who have breastfed previously, especially those with a history of mastitis.Riordan & wambach, 4th ed.P.294: breastfeeding and human lactation.However in this case, the mother¿s mastitis required prompt medical attention for pain relief and prescription antibiotics to avoid progression to overwhelming sepsis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PUMP IN STYLE RETAIL EN FR ES (YELLOWFIN)
Common Device Name
PUMP, BREAST, POWERED
Manufacturer (Section D)
MEDELA LLC
1101 corporate dr
mchenry IL 60050
Manufacturer (Section G)
MEDELA LLC
1101 corporate drive
mchenry IL 60050
Manufacturer Contact
dave kurudza
1101 corporate drive
mchenry, IL 60050
8004358316
MDR Report Key14999873
Report Number1419937-2022-00058
Device Sequence Number1614192
Product Code HGX
UDI-Device Identifier00020451401454
UDI-Public020451401454
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K200508
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number101041361
Device Catalogue Number101041361
Device Lot Number0000038770
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/01/2022
Distributor Facility Aware Date06/23/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/23/2022
Initial Report FDA Received Date07/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/21/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexFemale
-
-