• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDIA® FOLEY CATHETER INSERTION TRAY; FOLEY TRAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDIA® FOLEY CATHETER INSERTION TRAY; FOLEY TRAY Back to Search Results
Model Number 802130
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/23/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the syringe to fill balloon was broken when they tried to use it.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Additional Manufacturer Narrative
The reported event was inconclusive because no sample was returned.A potential root cause for this failure could be due to "poor interface of tip with inflation valve".The lot number was unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "directions for use: 1) remove tray lid 2) place under pad beneath patient, plastic side down.3) put on cuffed gloves 4) position drape on patient 5) open lubricant.Lubricate catheter 6) open packet of swabs 7) prep patient with swabs 8) proceed with catheterization in usual manner.To inflate catheter, simply insert tip of water filled syringe gently into valve (do not over penetrate) and depress plunger.Instill entire amount of sterile water.9) to deflate catheter balloon: gently insert a syringe in the catheter valve.Never use more force than is required to make the syringe stick in the valve.If you notice slow or no deflation reseat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If permitted by hospital protocol the valve arm may be severed.If this fails contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur care should be taken to assure that all balloon fragments have been removed from the patient.Warning : after use, the product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and applicable local, state and federal laws and regulations." the device was not returned.
 
Event or Problem Description
It was reported that the syringe to fill balloon was broken when they tried to use it.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDIA® FOLEY CATHETER INSERTION TRAY
Common Device Name
FOLEY TRAY
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
juan velez
8195 industrial blvd
covington 30014
7707846100
MDR Report Key15000007
Report Number1018233-2022-05388
Device Sequence Number19251045
Product Code OHR
UDI-Device Identifier00801741018275
UDI-Public(01)00801741018275
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number802130
Device Catalogue Number802130
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/23/2022
Supplement Date Received by Manufacturer10/12/2022
Initial Report FDA Received Date07/13/2022
Supplement Report FDA Received Date10/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-