It was reported that on (b)(6) 2022, the patient was hospitalized for pneumonia.During this admission, the patient's hemoglobin dropped and a large hematoma was observed in the patient's right upper limb.An x-ray was performed which ruled out a fracture of the limb.The patient received 4 units of packed red blood cells.
|
Manufacturer's investigation conclusion: a direct correlation between the reported bleeding and heartmate 3 left ventricular assist system (lvas), serial number (b)(6), cannot be conclusively determined through this evaluation.The relevant sections of the device history records were reviewed and showed no deviation from manufacturing or quality assurance specifications.Review of the sterilization and packaging documentation showed no deviation from manufacturing specifications.The heartmate 3 lvas instructions for use (ifu), rev.G, contains the following information: section 1 lists bleeding as an adverse event that may be associated with the use of heartmate 3 left ventricular assist system.No further information was provided.The manufacturer is closing the file on this event.
|