| Model Number |
82-6631 |
| Medical Device Problem Code |
Failure to Read Input Signal (1581)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/24/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
A facility reported a neuromonitor basic kit (id 826631) was placed on (b)(6) 2022.After the patient was sent back to ward the microsensor could not show a normal readings.They changed with a new icp monitor (id 826635) and new cable however, the microsensor still could not show normal reading with p-0 on the monitor and could not enter into the next step.Therefore, the microsensor was removed on (b)(6) 2022 and monitoring was stopped.
|
| |
|
Additional Manufacturer Narrative
|
|
The microsensor was returned for evaluation.Device history record (dhr) - the product 826631 for lot 218153 sn (b)(6) and the lot met specifications when released.Failure analysis - the he issue of the complaint has been confirmed.Connector wire broken inside.Excessive amount of foreign residue covering inside of connector.No testing was possible.The root cause could be determined as a mishandling of the device.Suspect fluid ingress the cause of the foreign residue inside of connector.
|
| |
|