Model Number A2079
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/23/2022
Type of Reportable Event Malfunction
Event or Problem Description
This is 1 of 2 reports linked to mfg report number 3004608878-2022-00144: a facility reported that during an unspecified surgery, when clamping the head, the locking screw on the mayfield swivel adapter that connects the mayfield infinity xr2 base unit (a2079) and the head mount broke.The patient's head was thus freely movable and was secured by the bedding orderly to avoid injuries.No serious consequences for the patient were observed.The event led to increased surgery time of 15 minutes.
 
Additional Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
Failure analysis - the a2079 mayfield base unit was not received for evaluation; however, the swivel adapter part 437a2400m, which is a component of the a2079, was returned and evaluated, but the entire base unit was not.Additionally, the broken screw was confirmed via customer provided pictures.Root cause - the complaint is confirmed via inspection of the unit.The screw of the swivel assembly was broken.Probable root cause includes excessive force and rough handling of the unit.No further investigation is required based on the acceptability of risk and no adverse trends were identified.This will be monitored and trended going forward.At present, we consider this complaint to be closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD INFINITY XR2 BASE UNIT
Common Device Name
MAYFIELD
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH
Manufacturer (Section G)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
cincinnati OH
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key15001205
Report Number3004608878-2022-00143
Device Sequence Number11206926
Product Code FWZ
UDI-Device Identifier10381780267959
UDI-Public10381780267959
Combination Product (Y/N)N
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberA2079
Device Catalogue NumberA2079
Device Lot Number16M0185
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 06/24/2022
Supplement Date Received by Manufacturer08/19/2022
Initial Report FDA Received Date07/13/2022
Supplement Report FDA Received Date08/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age68 YR
Patient SexMale
-
-