Model Number H7493941812250
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/24/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that stent dislodge occurred.A 2.50 x 12mm synergy xd drug-eluting stent was selected for use.However, upon loading the stent into the guidewire, the stent detached from the delivery balloon.The procedure was completed with a different device.There were no patient complications reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr: synergy xd mr us 2.50 x 12mm stent delivery system was returned for analysis.Examination of the crimped stent via scope found evidence of damage with stent struts from the proximal stent region lifted and pulled distally.The undamaged crimped stent outer diameter was measured and the result is within max crimped stent profile measurement.The balloon cones were reviewed via scope, and no issues were noted.The balloon wings were tightly wrapped and evenly folded.The balloon had not been subjected to positive pressure.A visual and microscopic examination of the bumper tip no issues.A visual and tactile examination of the hypotube shaft found multiple kinks along several locations of the hypotube shaft.A visual and tactile examination of the outer and mid-shaft section and a visual examination of the inner lumen found no issues.No other issues were identified during the product analysis.
 
Event or Problem Description
It was reported that stent dislodge occurred.A 2.50 x 12mm synergy xd drug-eluting stent was selected for use.However, upon loading the stent into the guidewire, the stent detached from the delivery balloon.The procedure was completed with a different device.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY XD
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15001718
Report Number2134265-2022-07663
Device Sequence Number11153701
Product Code NIQ
UDI-Device Identifier08714729980797
UDI-Public08714729980797
Combination Product (Y/N)Y
Initial Reporter StateMI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/09/2023
Device Model NumberH7493941812250
Device Lot Number0028452325
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/27/2022
Supplement Date Received by Manufacturer08/11/2022
Initial Report FDA Received Date07/13/2022
Supplement Report FDA Received Date08/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/09/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-