• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US 4.5 HEALIX ADVANCE AWL; BONE AWL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US 4.5 HEALIX ADVANCE AWL; BONE AWL Back to Search Results
Model Number 222314
Medical Device Problem Codes Material Deformation (2976); Illegible Information (4050)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Additional narrative: udi: (b)(4).The expiration date is currently unavailable.Initial reporter occupation: reporter is a j&j sales representative.Investigation summary: the product was returned to mitek for evaluation.Mitek then conducted visual inspection of device received provided by customer.Visual observation revealed that the sharp tip at the distal end appears to be slighly flattened.Also, scratches were found near the tip.Finally, the laser line was not visible for the correct insertion depth.As a potential cause cannot be associated to manufacturing, therefore a manufacturing record evaluation is not required.Based on the condition of the device, this complaint can be confirmed.It is determined that the reusable instrument is worn from repeated use and servicing, this failure can be attributed to normal field wear.For the tip condition could be related when the device might have been dropped or device was tapped against a hard surface accidentally.However, it cannot be conclusively affirmed.Lot number on the device indicates this device is 6 years old and hence this failure can be attributed to normal field wear.As per ifu: the precautions for this type of device consist of to inspect the condition of the device prior to use.This device can damage, worn or bent; therefore, when this occurs it need to replace.The instrument life is generally determined by the wear or damaged from handling or surgical use.At this point in time, no corrective action is required, and no further action is warranted.However, in depuy synthes mitek, additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.
 
Event or Problem Description
This is report 1 of 2 for (b)(4).It was reported by the sales rep that during a rotator cuff repair procedure on an unknown date, it was observed that the laser lines on the 4.5 healix advance awl device were very hard to see.During in-house engineering evaluation, it was determined that the sharp tip at the distal end on the device appeared to be slighly flattened.The procedure was completed successfully.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.5 HEALIX ADVANCE AWL
Common Device Name
BONE AWL
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key15001816
Report Number1221934-2022-02135
Device Sequence Number16588730
Product Code LXH
UDI-Device Identifier10886705022458
UDI-Public10886705022458
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number222314
Device Catalogue Number222314
Device Lot Number1612001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/20/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/01/2022
Initial Report FDA Received Date07/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/08/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-