MAUDE Adverse Event Report: CARDINAL HEALTH CURITY MONO-FLO 18FR WO PVPI; COLLECTOR, URINE, (AND ACCESSORIES) FOR INDWELLING CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDINAL HEALTH CURITY MONO-FLO 18FR WO PVPI; COLLECTOR, URINE, (AND ACCESSORIES) FOR INDWELLING CATHETER Back to Search Results
Model Number 6156MF
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.
 
Event or Problem Description
The customer reported that the urine accumulated at the connection between the running tube and the bag and did not flow.They drained the urine in the bag and pinched the chamber part of the connection between the bag and running tube, and although it may have flowed a little, they replaced it.There was no patient injury.
 
Additional Manufacturer Narrative
One decontaminated drainage bag was received at the manufacturing site for investigation.The date code printed on the bag corresponds with item code 6156mf and lot number 2108935464.The device history record was reviewed for lot 2108935464 and indicated that the product was released accomplishing all quality standards.The device was manufactured on april 08, 2021.Visual and functional inspections were performed, and the reported condition could not be confirmed; during the filling of the bag to its maximum capacity (2000 ml), no conditions were observed that prevented the flow of liquid through the pvc tube and the drip chamber into the bag.A gemba was carried out in the manufacturing process.The current process and controls were found to be followed correctly, including subassemblies, finished product assembly, packaging, and product inspections.No abnormal conditions were found that could trigger the reported condition.A corrective action is not applicable at this time.Functional testing and visual inspections are being performed according to our current quality standards and inspection procedures.However, the appropriate staff has been notified of the reported condition with the purposes of raising staff awareness.We will continue to monitor customer complaints and feedback notification for adverse trends that require immediate attention.This complaint will be used for tracking and trending purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CURITY MONO-FLO 18FR WO PVPI
Common Device Name
COLLECTOR, URINE, (AND ACCESSORIES) FOR INDWELLING CATHETER
Manufacturer (Section D)
CARDINAL HEALTH
777 west street
mansfield MA 02048
Manufacturer (Section G)
CARDINAL HEALTH
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX   22500
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183640
MDR Report Key15002206
Report Number9612030-2022-03316
Device Sequence Number1880596
Product Code KNX
UDI-Device Identifier10192253059809
UDI-Public10192253059809
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number6156MF
Device Catalogue Number6156MF
Device Lot Number2108935464
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/12/2022
Supplement Date Received by Manufacturer07/12/2022
Initial Report FDA Received Date07/13/2022
Supplement Report FDA Received Date08/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/08/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-