Manufacturer's investigation conclusion: review of the log files provided by the account confirmed low flow alarms.Although a specific cause for these events could not be conclusively determined through this evaluation, the account reported that the alarms were caused by intraoperative hypotension and worsening right ventricular (rv) failure.A direct correlation between heartmate (hm) 3 left ventricular assist system (lvas), serial number (b)(6), and the reported events and patient outcome could not be conclusively established.The controller event log file contained data from 30jun2022 through 08jul2022, per the timestamps.Low flow fault flags were captured on 08jul2022, when the estimated flow decreased below the low flow threshold of 2.5 liters per minute (lpm), to a range of 1.7-2.4 lpm.These faults resulted in 4 low flow hazard alarms, with the longest lasting over 57 minutes in duration.At 13:55:49 on 08jul2022, the driveline was disconnected from the system controller.This was followed by the disconnection of the black and white power cables and a controller shutdown sequence.These events were consistent with the reported controller exchange performed while the patient was in the or.The pump appeared to function as intended at the set speed while the driveline was connected to the system controller.It was also reported that heartmate 3 lvas, serial number (b)(6), would not be returned for evaluation.The relevant sections of the device history records for (b)(6) were reviewed and showed no deviations from manufacturing or quality assurance specifications.The implant kit was shipped on 15mar2020.The heartmate 3 lvas instructions for use (ifu) and the heartmate 3 patient handbook are currently available.Section 1 of the ifu, ¿introduction¿, lists potential adverse events, including right heart failure, that may be associated with the use of the heartmate 3 left ventricular assist system.Section 1 of the ifu also addresses all pump parameters, including pump flow.Section 4 of the ifu, ¿system monitor¿, describes the pump flow display and the hazard alarms.Per design, when the estimated flow value is calculated at less than 2.5 lpm, a low flow status is posted to the log file.If the flow remains below 2.5 lpm for 10 seconds, a low flow hazard alarm is triggered.Section 4 also explains that changes in patient condition can result in low flow.Section 6 of the ifu, ¿patient care and management¿ (under ¿right heart failure¿), states that patients may develop right ventricular failure during or shortly after implant and outlines the associated treatment options.Additionally, section 6, under ¿caution!¿, explains that right ventricular dysfunction, especially when combined with elevated pulmonary vascular resistance, may limit the effectiveness of the left ventricular assist system due to reduced filling of the pump.Section 7 of the ifu, ¿alarms and troubleshooting¿, and section 5 of the patient handbook, ¿alarms and troubleshooting¿, address all system alarm conditions as well as the appropriate actions associated with each condition.Furthermore, section 8 of the patient handbook, ¿handling emergencies¿, also provides examples of emergencies and the proper actions to take in the event an emergency occurs.No further information was provided.The manufacturer is closing the file on this event.
|