| Model Number |
D011 |
| Medical Device Problem Codes |
Premature Discharge of Battery (1057); Battery Problem (2885)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/06/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this device was suspected of exhibiting premature battery depletion behavior.The battery longevity decreased from 3 years to elective replacement indicator in approximately 4 months time.Subsequently, this device was explanted and replaced.No additional adverse patient effects were reported.This device is expected for return.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned inogen mini icd vr was analyzed and an engineering-level longevity prediction calculation was completed to assess the rate of battery depletion.Given the programmed parameters and other data stored within the memory of the device, the results of this calculation indicated that the actual rate of battery depletion fell within an acceptable range.Next, a series of diagnostic tests were conducted that verified the performance of pacing, sensing, defibrillation, and recording functions.Having met the engineering longevity prediction, functionally passed all returned product testing, and with no further information to indicate a product performance issue, we have concluded that this device experienced normal battery depletion.
|
| |
|
Event or Problem Description
|
|
It was reported that this device was suspected of exhibiting premature battery depletion behavior.The battery longevity decreased from 3 years to elective replacement indicator in approximately 4 months time.Subsequently, this device was explanted and replaced.No additional adverse patient effects were reported.This device was received for analysis.
|
| |
|
Search Alerts/Recalls
|