• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. TRILOGY 100; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. TRILOGY 100; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number 1054260
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/06/2022
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer received information alleging visualization of particles in the air path.There was no report of serious patient harm or injury.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
 
Additional Manufacturer Narrative
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer previously received information alleging visualization of particles in the air path related to a ventilator's sound abatement foam.Repeated attempts to have the device and components returned for evaluation and investigation were unsuccessful.The manufacturer believes they will be unable to gather additional information.The manufacturer is submitting an updated report at this time.If pertinent information becomes available to the manufacturer at a later date, a follow-up report will be filed.Section h6 updated in this report.
 
Additional Manufacturer Narrative
Additional information was received on jan 24, 2025 and section h11 should be reported as: the manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer previously received information alleging visualization of particles in the air path related to a ventilator's sound abatement foam.The device was returned to the manufacturer's service center for further evaluation.The device was evaluated.There was no mention of visual findings to the external part of the device.The internal aspect of the device was inspected.The device powered on and airflow was confirmed.The device's downloaded logs were reviewed.There were no errors found.The manufacturer's service center concludes that they confirm the customer's allegation and there was a visible foam degradation.Device was scrapped.Sections d8, d9, h2, h3 and h6 has been updated in this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY 100
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
rosko melissa
6501 living place
6501 living place
pittsburgh, PA 15206
7243512041
MDR Report Key15004096
Report Number2518422-2022-60200
Device Sequence Number1880303
Product Code CBK
UDI-Device Identifier00606959028944
UDI-Public00606959028944
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K083526
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/24/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number1054260
Device Catalogue Number1054260
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/06/2022
Supplement Date Received by Manufacturer07/27/2023
01/24/2025
Initial Report FDA Received Date07/13/2022
Supplement Report FDA Received Date07/28/2023
07/24/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/14/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
FDA RES 88071
Patient Sequence Number1
Patient SexUnknown
-
-