• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWERPICC 5F SINGLE-LUMEN POLYURETHANE CATHETER RADIOLOGY BASIC TRAY; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWERPICC 5F SINGLE-LUMEN POLYURETHANE CATHETER RADIOLOGY BASIC TRAY; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number 3175335
Medical Device Problem Codes Fluid/Blood Leak (1250); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/22/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The date of event was not provided by the complainant/reporter, the date reflected in this report is the date bd became aware of the event.The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of refy3552 showed five other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported by the customer leaking picc.It was stated the picc was checked and it didn't seem to be cracked.It's like there is a problem with the threading.
 
Event or Problem Description
It was reported by the customer leaking picc.It was stated the picc was checked and it didn't seem to be cracked.It's like there is a problem with the threading.Additional information received: picc line is leaking between the end of the picc and the clave.¿ what type of catheter securement was utilized? stat-lok.¿ date of occurrence: jun 22/22¿ placement date: jun 14/22.¿ explant date: picc still in place.¿ was there any patient harm reported: no.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample analysis, manufacture quality and production records, and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a leaking catheter was confirmed, and the cause was manufacturing-related.The product returned for evaluation was one 5fr s/l powerpicc catheter.Light usage residues were observed throughout the sample and a needleless injection cap was attached to the luer adapter.Microscopic inspection of the luer adapter revealed a protrusion in the luer material within the orifice.That material had the appearance of a mold defect which could result in inadequate sealing of the luer hub with a syringe.Bd is working closely with the manufacture facility to prevent recurrence of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPICC 5F SINGLE-LUMEN POLYURETHANE CATHETER RADIOLOGY BASIC TRAY
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
johanna de oliveira
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key15004583
Report Number3006260740-2022-02678
Device Sequence Number13897364
Product Code LJS
UDI-Device Identifier00801741027697
UDI-Public(01)00801741027697
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K033389
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number3175335
Device Catalogue Number3175335
Device Lot NumberREFY3552
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/22/2022
Supplement Date Received by Manufacturer08/31/2022
Initial Report FDA Received Date07/13/2022
Supplement Report FDA Received Date09/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-