• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGELINI THERMACARE LOWER BACK & HIP 8HR L/XL 2CT; HOT OR COLD DISPOSABLE PACK.

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGELINI THERMACARE LOWER BACK & HIP 8HR L/XL 2CT; HOT OR COLD DISPOSABLE PACK. Back to Search Results
Model Number ThermaCare HeatWraps - Lower Back & Hip
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Burn(s) (1757); Burning Sensation (2146); Ulcer (2274)
Date of Event 06/26/2022
Type of Reportable Event Serious Injury
Event or Problem Description
On (b)(6) 2022, a spontaneous report from the united states was received via telephone from a 77-year-old male who used a thermacare lower back and hip l/xl 8hr heat wrap.Medical history include pain.Further medical history and concomitant products were not provided.On (b)(6) 2022, the consumer topically applied a thermacare lower back and hip l/xl 8hr heat wrap to his lower back.On (b)(6) 2022, after using the product, the consumer took a shower and felt a burning sensation from a dime size area on his lower back.The area looked like a burn.The heat wrap was intact and without any tears.On (b)(6) 2022, the consumer called a doctor and a nurse advised him to apply aloe vera cream to the affected area.As of (b)(6) 2022, the consumer anticipated making an appointment with her physician.No additional information was provided.
 
Additional Manufacturer Narrative
The root cause cannot be identified.There is limited device specific information provided, no batch number or return sample was available for evaluation.Without a batch reference number or return sample, a manufacturing and technical evaluation cannot be completed for the wrap involved in this case.No product quality related trend was identified for the subclass adverse event safety request for investigation.Care should be taken when using the device, following all safety and use information as provided with the wrap to avoid the risk of blisters and other skin irritations.The manufacturing operations employ quality control procedures which include in process testing, thermal testing and visual inspection, to ensure packaged product quality.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMACARE LOWER BACK & HIP 8HR L/XL 2CT
Common Device Name
HOT OR COLD DISPOSABLE PACK.
Manufacturer (Section D)
ANGELINI
1231 wyandotte dr
albany GA 31705
Manufacturer (Section G)
BRIDGES CONSUMER HEALTHCARE
811 broad street, suite 600
chattanooga TN 37402
Manufacturer Contact
scott hughes
811 broad street, suite 600
chattanooga, TN 37402
4237178579
MDR Report Key15004792
Report Number3007593958-2022-00043
Device Sequence Number1562359
Product Code IMD
UDI-Device Identifier00305733010037
UDI-Public00305733010037
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberThermaCare HeatWraps - Lower Back & Hip
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/29/2022
Initial Report FDA Received Date07/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age77 YR
Patient SexMale
Patient Weight73 KG
-
-