| Model Number |
N/A |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/23/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of regp2539 showed no other similar product complaint(s) from this lot number.
|
| |
|
Event or Problem Description
|
|
It was reported by healthcare professional "i placed a picc yesterday and was called back to facility today due to no blood return and leaking.I noted the picc was leaking at the connection between the hub and the extension tubing.The picc was removed." additional information received 06/30/2022: what type of catheter securement was utilized? statlock date of occurrence: (b)(6) 2022 placement date: (b)(6) 2022 explant date: (b)(6) 2022.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, trend analysis, applicable previous investigation(s), sample (if available), and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a leaking catheter was confirmed.The product returned for evaluation was one 4fr s/l powerpicc.The investigation findings are consistent with catheter damage caused by contact with a sharp edged instrument such as a scalpel.The returned product sample was evaluated and a split was observed in the extension tubing, near the molded joint.Microscopic examination of the catheter split confirmed it was typical of contact with a sharp bladed instrument, and the characteristics observed which supported this type of failure included: ¿ the fracture surface was reflective in nature and contained a striated pattern.This can occur due to pattern transfer of the sharpened edge typically found on a scalpel-type instrument.¿ sharply formed fracture edges ¿ planar shape to the fracture surface an examination of the catheter structure revealed no potential damage/defect related to manufacture of the product.H3 other text : evaluation findings are in section h.11.
|
| |
|
Event or Problem Description
|
|
It was reported by healthcare professional "i placed a picc yesterday and was called back to facility today due to no blood return and leaking.I noted the picc was leaking at the connection between the hub and the extension tubing.The picc was removed." additional information received 06/30/2022: ¿ what type of catheter securement was utilized? statlock ¿ date of occurrence: (b)(6) 2022 ¿ placement date: (b)(6) 2022 ¿ explant date: (b)(6) 2022.
|
| |
|