| Model Number |
N/A |
| Medical Device Problem Code |
False Positive Result (1227)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/22/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An (b)(6) customer observed a reactive immulite 2000 xpi toxoplasma quantitative igg (txp) result for a patient sample that was nonreactive on an alternate method.The instrument was checked for the following by field servce engineer: dispensing test, sample and reagent probes calibration, washing and substrate dispensing check, alignments and dilution speed station test, water test.The limitations section of the instructions for use (ifu) states, "the results of the test must be taken within the context of the patient's clinical history, symptomology and other laboratory findings." siemens is investigating.
|
| |
|
Event or Problem Description
|
|
The customer observed a reactive immulite 2000 xpi toxoplasma quantitative igg (txp) result for a patient sample that was nonreactive on an alternate method.The initial result was reported to the physician(s) who questioned the result.There are no reports that treatment was altered or prescribed or adverse health consequences due to the reactive txp result.
|
| |
|
Additional Manufacturer Narrative
|
|
The initial mdr 1219913-2022-00217 was filed on july 13, 2022.An outside the united states customer observed a reactive immulite 2000 xpi toxoplasma quantitative igg (txp) result for a patient sample that was nonreactive on an alternate method.Additional information - august 30, 2022: it is unknown whether the reactive sample was tested with hbt/ nabt or with any other lot of immulite 2000 txp.Siemens is investigating.
|
| |
|
Additional Manufacturer Narrative
|
|
The initial mdr was filed on july 13, 2022 and mdr supplemental 1 was filed on september 19, 2022 and supplemental 2 was filed on november 7, 2022.An outside the united states customer observed a reactive immulite 2000 xpi toxoplasma quantitative igg (txp) result for a patient sample.The sample was nonreactive on an alternate method.Additional information - november 14, 2022: the customer has been unable to run any samples on an alternate immulite.The customer confirmed that they do not have enough sample volume to send any samples to siemens for testing.The customer confirms that they send the samples to another lab.Siemens is investigating.
|
| |
|
Additional Manufacturer Narrative
|
|
The initial mdr 1219913-2022-00217 was filed on july 13, 2022, mdr 1219913-2022-00217 supplemental 1 was filed on september 19, 2022, mdr 1219913-2022-00217 supplemental 2 was filed on november 7, 2022 and mdr 1219913-2022-00217 supplemental 3 was filed on december 5, 2022 and mdr 1219913-2022-00217 supplemental 4 was filed on january 12, 2023.An outside the united states customer observed a reactive immulite 2000 xpi toxoplasma quantitative igg (txp) result for a patient sample.The sample was reactive upon repeat testing on the same system but nonreactive on an alternate method.Additional information - february 21, 2023: a customer reported an increase of false reactive results since the use of immulite 2000 toxoplasma quantitative igg (txp) lot 519.The samples were repeated on immulite 2000 xpi and still recovered reactive.They sent the samples to a different laboratory and the results were non-reactive when tested with an alternate method.Siemens reviewed the customer's adjustment and qc data and found them to be valid and comparable with release.Further investigation by siemens was performed.The investigation aimed to determine if there had been a positive shift with kit lots 517, 519 or 520 and used internal and smart remote services (srs) data to assess this.Smart remote services (srs) data was gathered and the percentage of samples resulting reactive, indeterminate and non-reactive for each kit lot was calculated.An increase in %reactive results is not observed with all lots.Srs data and internal data do not confirm a positive shift in performance of immulite 2000 toxoplasma kit lots 517, 519 and 520.Based on panel data interpretation, variability for a single sample is observed historically.Review of the precision of the assay using kit lots 517 and 519 showed the precision to be within specifications.The investigation confirmed there has been no positive shift in performance for kit lots 517, 519 or 520.Siemens cannot rule out pre-analytical factors such as sample handling or integrity, or sample specific effects.A field service engineer checked the system and verified its performance.The spyfile provided does not show any abnormalities.Evaluation of the errorlog did not indicate any recurring errors that would indicate an instrument malfunction or error condition contributed to questionable results.Based on review of data available, it is inconclusive to determine whether an instrument issue contributed to the questionable sample results.
|
| |
|
Additional Manufacturer Narrative
|
|
The initial mdr 1219913-2022-00217 was filed on july 13, 2022 and mdr 1219913-2022-00217 supplemental 1 was filed on september 19, 2022 and 1219913-2022-00217 supplemental 2 was filed on november 7, 2022 and mdr 1219913-2022-00217 supplemental 4 was filed on december 5, 2022.An outside the united states customer observed a reactive immulite 2000 xpi toxoplasma quantitative igg (txp) result for a patient sample.The sample was nonreactive on an alternate method.Additional information - december 22, 2022: siemens reviewed the customer's adjustment and qc data and found them to be valid and comparable with release.A field service engineer checked the system and verified its performance.Siemens reviewed patient field data and found that lot: 519 interpretation recovery was comparable to other lots.Siemens initiated a study with txp lots: 517, 519 and 520 with qc and patient samples, but the data was inconclusive.A second study was performed with txp lots: 517, 520 and 521 (519 lot had expired) with qc and a patient sample panel and noted high precision on some samples and different interpretations between lots.The customer used txp lot: 520 and found recovery to be better.Siemens is continuing to investigate.
|
| |
|
Additional Manufacturer Narrative
|
|
The initial mdr 1219913-2022-00217 was filed on july 13, 2022 and mdr 1219913-2022-00217 supplemental 1 was filed on september 19, 2022.An outside the united states customer observed a reactive immulite 2000 xpi toxoplasma quantitative igg (txp) result for a patient sample.The sample was nonreactive on an alternate method.Additional information - october 13, 2022: the spyfile provided shows 2 samples that were run on may 31, 2022:(b)(6).Both samples do not show any abnormalities as far as sample or reagent level sensing is concerned.In addition sample level sensing within the dilution well appeared to be normal.Evaluation of the errorlog on may 31, 2022 did not indicate any recurring errors that would indicate an instrument malfunction or error condition contributed to questionable results.Based on review of data available it is inconclusive to determine whether an instrument issue contributed to the questionable sample results.Spyfiles and errorlogs have not been provided for the sample tested on june 22, 2022.Siemens is investigating.
|
| |
|
Search Alerts/Recalls
|