• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL GRYPHON SLOTTED SAWTOOTH GUIDE; BONENAIL GUIDEWIRE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL GRYPHON SLOTTED SAWTOOTH GUIDE; BONENAIL GUIDEWIRE Back to Search Results
Model Number 210824
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).The expiration date and device manufacture date were unknown.The initial reporter is a j&j sales representative.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event or Problem Description
It was reported by sales rep that during a labrum repair on (b)(6) 2022, it was observed that the gryphon slotted sawtooth guide device had separation of the shaft.Another like device was used to complete the procedure.There were no adverse patient consequences nor surgical delay reported.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: the device was received and evaluated, on visual inspection, it could be observed that the shaft is separated from the handle due to welding detachment.As a potential cause cannot be associated to manufacturing, therefore a manufacturing record evaluation is not required.As part of depuy synthes mitek quality process all devices are manufactured, inspected, and released to approved specifications.According with the visual inspection, this complaint can be confirmed.The possible root cause can be attributed to the rough and continuous use of the device.Since this is a reusable device, the constant manipulation between surgeries and sterilization process can lead to damages to the instrument.As per the instructions for use; end of useful instrument life is generally determined by wear or damage from handling or surgical use.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: d9, h3, h6: the actual device has been returned and is currently pending evaluation.Once the device has been evaluated, a supplemental medwatch report will be sent accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GRYPHON SLOTTED SAWTOOTH GUIDE
Common Device Name
BONENAIL GUIDEWIRE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin blanc 38
le locle CH-24 00
SZ  CH-2400
Manufacturer (Section G)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle CH-24 00
SZ   CH-2400
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key15007300
Report Number1221934-2022-02158
Device Sequence Number17856795
Product Code LXH
UDI-Device Identifier10886705024063
UDI-Public10886705024063
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number210824
Device Catalogue Number210824
Device Lot Number1607002
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/01/2022
Supplement Date Received by Manufacturer07/19/2022
08/14/2022
Initial Report FDA Received Date07/13/2022
Supplement Report FDA Received Date07/20/2022
08/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-