• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT GMBH ARCHITECT SYPHILIS TP REAGENT KIT ENZYME LINKED IMMUNOABSORPTION ASSAY, TREPONEMA PALLIDUM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT GMBH ARCHITECT SYPHILIS TP REAGENT KIT ENZYME LINKED IMMUNOABSORPTION ASSAY, TREPONEMA PALLIDUM Back to Search Results
Catalog Number 08D06-77
Device Problem False Negative Result (1225)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/29/2022
Event Type  malfunction  
Event Description
The customer stated that nonreactive architect syphilis results of 0. 59 and 0. 66 s/co were generated for a 47 year old female patient in the dermatology clinic diagnosed with psoriasis vulgaris. Tppa testing was positive. The customer suspects the architect syphilis results to be false nonreactive. No impact to patient management was reported.
 
Manufacturer Narrative
An evaluation is in process. A follow-up report will be submitted when the evaluation is complete. All available patient information was included. Additional patient details are not available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameARCHITECT SYPHILIS TP REAGENT KIT
Type of DeviceENZYME LINKED IMMUNOABSORPTION ASSAY, TREPONEMA PALLIDUM
Manufacturer (Section D)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM 65205
Manufacturer (Section G)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM 65205
Manufacturer Contact
siobhan wright
lisnamuck
post market surveillance
longford N39 E-932
EI   N39 E932
433331157
MDR Report Key15007528
MDR Text Key297053160
Report Number3002809144-2022-00242
Device Sequence Number1
Product Code LIP
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K153730
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 07/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/13/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number08D06-77
Device Lot Number32478BE00
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/25/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/22/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No

Patient Treatment Data
Date Received: 07/13/2022 Patient Sequence Number: 1
Treatment
ARC I2000SR INST, 03M74-02, ISR54017; ARC I2000SR INST, 03M74-02, ISR54017
-
-