| Model Number |
A2020-120 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/22/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the procedure was performed to treat a lesion in the heavily calcified peroneal artery, via antegrade approach.The armada 14 balloon dilatation catheter was advanced to the leion without difficulty.The balloon was inflated 1 time to 12 atmospheres (atm) and a balloon rupture occurred.The device was removed without difficulty and no adverse patient effect occurred.There was no adverse patient effect and no clinically significant delay in the procedure.A second same device was used without issue.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation determined the reported balloon rupture appears to be related to operational context.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|
| |
|