| Model Number |
MER3700 |
| Medical Device Problem Code |
Incorrect Measurement (1383)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/24/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the programmer exhibited an incorrect longevity estimate because of a battery longevity calculation overestimation.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted for the programmer as a part of the programmer longevity estimator software error field action (fda recall number: fa-q222-crm1) set of complaints.Should any additional information be obtained, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
The longevity estimation anomaly near elective replacement indicator (eri) was confirmed by reviewing the data in the device image and session records.The longevity overestimation was determined to be due to an estimation inaccuracy in the merlin programmer remaining longevity calculation.
|
| |
|