Reference initial mdr report # 9680001-2022-00107.On (b)(6) 2022 our distributor provided three possible lot numbers for the involved device and requested a device history record review for each lot.All lot numbers provided indicate (b)(6) usa as the place of manufacture and not (b)(6) as originally reported.This report is being filed to correct the place of manufacture for this event.The device was not received for evaluation at the time of this report; therefore, no physical analysis of the device can be performed.The actual lot number of the product used in the event is unknown; however, the distributor provided three possible lot numbers on (b)(6) 2022: lot numbers 6229663; 6229662; 6146673.All three lot numbers identified the same device model as indicated in section d4.A review of the device history records of the supplied lot numbers for the specimen device does not present any indication of manufacturing defect or anomaly that could have impacted on the event as reported.During manufacturing the production operators are instructed to 100% visually and tactilely inspect for any obvious defect, which includes a tactile examination of the entire length of each wire.In addition, during packaging of this product the operators are instructed to 100% visually inspect for any obvious defect prior to shipment.The history records indicate this product was final inspection tested at lake region medical and was determined to be acceptable.As noted in the device instructions for use (dfu) warnings: manipulate the zipwire hydrophilic guidewire slowly and carefully in the urinary system while confirming the behavior and location of the wire's tip under fluoroscopy.Excessive manipulation of the zipwire hydrophilic guidewire without fluoroscopic confirmation may result in perforation or trauma of the linings or associated tissues, channels or ducts.If any resistance is felt or if the tip's behavior and/or location seem improper, stop manipulating the wire and/or catheter and determine the cause by fluoroscopy.Failure to exercise proper caution may result in bending, kinking, separation of the zipwire hydrophilic guidewire's tip, damage to the catheter, or damage to the urinary system.If necessary, remove the zipwire hydrophilic guidewire and ancillary device or scope as a complete unit to avoid complications.Use extreme caution when using a laser, making sure to avoid contact with the zipwire hydrophilic guidewire.Direct contact may cause damage to the wire and/or sever the wire.The dfu precautions also indicate: the zipwire hydrophilic guidewire should be advanced through the scope using short, deliberate 2-3cm movements to prevent inadvertent damage to the device or patient.Due to variations in certain catheter tip diameters, abrasion of the hydrophilic coating may occur during manipulation.If any resistance is felt during introduction of the catheter, it is advisable to stop using such catheters.At this time, it is not possible to assign a definitive root cause for the event as reported.Based on the information provided, it appears that clinical and/or procedural factors impacted on the event as reported.If there is any further relevant information provided, a follow up medwatch report will be submitted.
|