EDWARDS LIFESCIENCES, PR FOGARTY THRU-LUMEN EMBOLECTOMY CATHETER; CATHETER, EMBOLECTOMY Back to Search Results
Model Number 12TLW804F
Medical Device Problem Codes Material Disintegration (1177); Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/27/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
One 12tlw804f fogarty embolectomy catheter with attached three way stopcock bd 3 ml syringe was returned for evaluation.The customers report of the balloon not inflating was confirmed.The balloon latex appeared discolored and deteriorated.The balloon did not inflate due to leakage from distal side of the balloon latex.A maze of cracks and holes were observed on the balloon.After cutting the proximal windings, the edges of the holes and cracks did not appear to match up.Latex deterioration is a condition usually caused by age, excessive exposure to light, atmosphere, or ozone.Appears on the balloon surface as a maze of fine cracks or crazing.The condition may occur in a small area or cover the entire balloon.There was no visible damage observed from both windings and the catheter body.An engineering evaluation was assigned to the manufacturing site for further investigation.Complaint histories for all reported events are reviewed against trending control limits monthly, and any excursions above the control limits are assessed and documented as part of this monthly review.
 
Event or Problem Description
It was reported that before use, the balloon would not inflate during the inflation test.The device was replaced, and the problem was solved.There were no patient complications reported.The product was returned for evaluation.The balloon latex appeared discolored and deteriorated.The edges of the holes and cracks did not appear to match up.
 
Additional Manufacturer Narrative
A device history record review was completed and documented that the device met all specifications upon distribution.Complaint histories for all reported events are reviewed against trending control limits, and any excursions above the control limits are assessed and documented as part of this monthly review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOGARTY THRU-LUMEN EMBOLECTOMY CATHETER
Common Device Name
CATHETER, EMBOLECTOMY
Manufacturer (Section D)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco CA 00610
Manufacturer (Section G)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer Contact
jessica atallah
1 edwards way
irvine, CA 92614
9492500294
MDR Report Key15008354
Report Number2015691-2022-06760
Device Sequence Number6753074
Product Code DXE
UDI-Device Identifier00690103043518
UDI-Public(01)00690103043518(17)220526(11)200225(10)62939801
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K892410
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/26/2022
Device Model Number12TLW804F
Device Catalogue Number12TLW804F
Device Lot Number62939801
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/16/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/27/2022
Supplement Date Received by Manufacturer07/28/2022
Initial Report FDA Received Date07/13/2022
Supplement Report FDA Received Date08/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/29/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-