Model Number 623-00-36F
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code Bone Fracture(s) (1870)
Date of Event 06/17/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Device not returned to the manufacturer.
 
Event or Problem Description
It was reported through the submission of a revision implant sheet that the patient's right hip was revised.Per sales rep: "patient had a periprosthetic fracture.Stem was revised and replaced with a longer monolithic stem.Shell stayed intact and liner was revised to an mdm." comparing records from patient's prior revision indicate that a size 7 anato stem, 36 +5 biolox head, and 36f 0° liner were revised.Update: there was a fracture laterally and slightly distal from the greater troch.Minor wear on the liner and head was observed.
 
Additional Manufacturer Narrative
Reported event: an event regarding wear involving a trident liner was reported.The event was not confirmed.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: no medical records were provided for review with clinical consultant.Product history review: review of the device history records indicate the devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to periprosthetic fracture and wear of the liner and head.The exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the device, pathology reports, pre- and post-operative x-rays and the primary/revision operative reports as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
It was reported through the submission of a revision implant sheet that the patient's right hip was revised.Per sales rep: "patient had a periprosthetic fracture.Stem was revised and replaced with a longer monolithic stem.Shell stayed intact and liner was revised to an mdm." comparing records from patient's prior revision indicate that a size 7 anato stem, 36 +5 biolox head, and 36f 0° liner were revised.Update: "there was a fracture laterally and slightly distal from the greater troch.Minor wear on the liner and head was observed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIDENT 0° X3 INSERT 36MM ID
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
anna ryan
raheen business park
limerick, NJ NA
EI   NA
2018315000
MDR Report Key15008458
Report Number0002249697-2022-01004
Device Sequence Number6819705
Product Code LPH
UDI-Device Identifier07613327039535
UDI-Public07613327039535
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K033716
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2015
Device Model Number623-00-36F
Device Catalogue Number623-00-36F
Device Lot NumberMJEW50
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/17/2022
Supplement Date Received by Manufacturer09/16/2022
Initial Report FDA Received Date07/13/2022
Supplement Report FDA Received Date10/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/06/2010
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age89 YR
Patient SexMale
Patient Weight91 KG
-
-