Summary of event: as reported, a small piece of the distal tip of a cxi support catheter separated prior to being used in the patient.The user had difficulty advancing the device onto the wire when they noticed the separated tip.The device did not make patient contact.The procedure was successfully completed with a replacement device.No unintended section of the device remained inside of the patient's body.No additional procedures were required due to the occurrence.No adverse effects were reported due to the occurrence.Investigation evaluation: reviews of the complaint history, device history record, instructions for use (ifu), manufacturing instructions, and quality control procedures were conducted during the investigation.A visual inspection of the complaint device was also conducted.The complaint device was returned to cook for investigation.The separated section of the tip was four millimeters, and the portion still attached to the catheter was five millimeters.A document-based investigation evaluation was performed.No related non-conformances were found, and there have been no other reported complaints for this lot number.The product ifu states, ¿upon removal from package, inspect product to ensure no damage has occurred.¿ the information provided upon review of the dmr, dhr, ifu, and investigation of the returned device suggests that there is evidence the device was manufactured out of specification; however, there is no evidence of additional non-conforming devices in-house or in the field.Cook has concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.Based on the information provided and the results of the investigation, cook has concluded that a manufacturing/quality control deficiency contributed to this incident.The risk analysis for this failure mode was reviewed and no additional escalation was required.The appropriate personnel have been notified and cook will continue to monitor for similar events.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|