| Model Number |
PT100 |
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/16/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The subject device is no longer in use by the healthcare facility.(b)(6) has requested for the complaint device to be returned for evaluation.We will provide a follow up report upon completion of our investigation.The pt100 myairvo 2 humidifier device is an active humidifier with integrated flow generator that delivers high flow warmed and humidified respiratory gases to spontaneously breathing patients through a variety of patient interfaces.It's intended use is for the treatment of spontaneously breathing patients who would benefit from receiving high flow warmed and humidified respiratory gases.The myairvo 2 device should not be used for life support purposes, and appropriate patient monitoring must be used at all times.
|
| |
|
Event or Problem Description
|
|
A distributor in florida reported that a pt100 myairvo 2 humidifier did not have an audible alarm.There was no patient involvement as the issue was found whilst the device was not in use on a patient.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4), sections d1, d2 and d4 were corrected.The brand name and common device name (d1 and d2) were changed from airvo 2 humidifier to myairvo2 humidifier.Model number and catalog number (d4) were changed from pt101 to pt100.The pt100 myairvo 2 humidifier device is an active humidifier with integrated flow generator that delivers high flow warmed and humidified respiratory gases to spontaneously breathing patients through a variety of patient interfaces.It's intended use is for the treatment of spontaneously breathing patients who would benefit from receiving high flow warmed and humidified respiratory gases.The myairvo 2 device should not be used for life support purposes, and appropriate patient monitoring must be used at all times.Method: the complaint myairvo 2 was received by fisher & paykel healthcare (f&p) where it was inspected by a trained f&p employee.The device was performance tested and the audible alarm function was checked.Results: performance testing revealed that the audible alarm of the myairvo 2 unit was working.Conclusion: we are unable to determine what may have caused the reported failure as there was no fault found with the returned device.
|
| |
|
Event or Problem Description
|
|
A distributor in florida reported that a pt100 myairvo 2 humidifier did not have an audible alarm.There was no patient involvement as the issue was found whilst the device was not in use on a patient.
|
| |
|