MEDTRONIC PUERTO RICO OPERATIONS CO. INTELLIS; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR
|
Back to Search Results |
|
Model Number 97715 |
Device Problems
High impedance (1291); Application Program Problem (2880); Communication or Transmission Problem (2896); Adverse Event Without Identified Device or Use Problem (2993); Patient Device Interaction Problem (4001)
|
Patient Problems
Inadequate Pain Relief (2388); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/22/2021 |
Event Type
Injury
|
Event Description
|
Information was received from a patient who was implanted with an implantable neurostimulator (ins) for spinal pain indications.The reason for call was 'settings not available'.Pt reported since implanted pt had had a lot of issues with therapy not working.Pt had seen a manufacturing representative (rep) a lot and last time rep set it on a.Pt said they were getting maybe 20% relief which was better than nothing.About a week ago pt started getting 'settings not available'.Pt said at first the device did not seem to want to recognize the ins.Pt was having to hold it right no top of ins before it would recognize it and let pt charge it.Pt confirmed ins was charged now but 'settings not available' was not resolved.Pt confirmed they had no falls/no trauma.Instructed pt to try other groups.Pt used the controller and had only group a.Additional information was reported that a manufacturing representative (rep) called in today and was with the pt.The caller noted she ran a 'connectivity check' and all contacts were red.The caller went on to run an impedance check and noted that all contacts are showing 'red' and >40k ohms.The caller states that the pt has been swimming a lot but noted there has been no falls/trauma.Technical services (tss) reviewed this is the reason for the 'settings not available' and reviewed that considering there are no viable impedance pairs the pt is likely not going to have therapy.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that patient¿s physician ordered x-rays of the device to check for any possible changes from initial implant.And to also check for any disconnections in the ipg.X-rays did not show any disconnections or changes from initial implant.Patient is being referred to his implanting physician for further evaluation.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received from the manufacturer representative.It was reported that they opened the ins pocket and tested the lead with a wens and most of the contacts were high impedance.Rep said the dr wanted to change the ins, so the dr used another ins and connected it to the lead and most of the contacts were still high impedance.Rep reports the dr is going to replace the lead.Rep called technical services (ts) after the dr replaced the ins.Ts reviewed that testing with the wens indicated the issue is with the lead.
|
|
Manufacturer Narrative
|
Concomitant medical products: product id 977c165 lot#/serial# (b)(6) product type lead.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
H1: corrected and checked serious injury.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
H3: product id# 97715, serial# (b)(6) was returned for analysis.Analysis found no significant anomalies and the implantable neurostimulator (ins) passed functional testing.Product id 977c165, serial# (b)(6) was returned for analysis.Analysis identified that all the conductors on the lead leg for circuits 0-7 were broken 21.5 cm from the distal end.All conductors on the lead leg for circuits 8-15 were broken 23.2 cm from the distal/proximal end of the lead.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|