BIOSENSE WEBSTER INC THERMOCOOL SF CARTOXP,B-D,TC; CARDIAC ABLATION PERCUTANEOUS CATHETER Back to Search Results
Model Number BNI35BDH
Medical Device Problem Codes Device Contamination with Body Fluid (2317); High Readings (2459); Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Cardiac Tamponade (2226)
Date of Event 06/15/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a male patient underwent a premature ventricular contraction (pvc) ablation procedure with two thermocool® sf nav bi-directional catheters (2) and suffered cardiac tamponade requiring pericardiocentesis.During the pvc procedure, the catheter was advanced to the target site right ventricle (rv), impedance was shown as about 200 ohm.They exchanged to a new cable but there is no change on the impedance, and then, the catheter was replaced with a new catheter of the same type but the value did not change.Therefore, when the catheter was moved to another position, the value was around 130 ~ 180 ohm.First ablation was conducted as 30w/5sec and impedance was 53 ohm down from the starting of 234 ohm.The second ablation was conducted at 30w/5sec and impedance was 52 down from the starting of 204 ohm.After switching to lifeline ablaze, decreased blood pressure was confirmed.Then, setting was set at 50 w with ablaze (w was changed at 1 ~ 40 w) and ablation was started at 135 ohm (max 152 ohm), 27 seconds ablation was performed.Cardiac tamponade was confirmed by echo, and drainage was started.Echo was confirmed again, blood pressure returned to normal limits, and the patient returned to the room.The physician's opinion on the relationship between the event and the product is that there may have been an impedance abnormality.The adverse event occurred on (b)(6) 2022.It was discovered post use of biosense webster catheters.The two defective bwi catheters were removed due to product malfunction, the ablation procedure was continued with another company's catheter and adverse event was confirmed when the ablation was performed with the other company's ablation catheter.The physician¿s opinion on the cause of this adverse event is that it was due to the procedure.The patient outcome of the adverse event is improved.Prior to noting the pericardial effusion, ablation was performed.There was no evidence of a steam pop.The event occurred during the ablation phase.The ablation never continued beyond the cut-off value.Generator parameters was power control mode.The complaint products will be returned for analysis.No error was message observed during the procedure.No issue related to the temperature and flow on catheter was observed.A smartablate generator with product code m4900207 and serial number was (b)(4) was used.The catheter will be returned for analysis.No neurological symptoms were observed on the patient.The duration of ablation used was less than 60 seconds.This complaint will be coded with char.Follow up has been done to clarify the response received that no char /coagulum/thrombus/clot was confirmed as there is discrepancy with the photograph provided.Should additional information become available this complaint will be reassessed.Since the event (cardiac tamponade) is life threatening and required intervention and prolonged hospitalization to prevent permanent impairment of a body function or permanent damage to a body structure, is to be considered serious and mdr-reportable.The high impedance was assessed as not mdr reportable.Since the user-defined cut-off was not exceeded the potential that it could cause or contribute to a death, serious injury, or other significant adverse event, is remote.Char was assessed as not mdr reportable.Char is a physical phenomenon of radiofrequency (rf) energy delivery and can be the normal result of the ablation process.The presence of char on the electrodes does not represent a serious injury in itself, nor is it necessarily the result of device malfunction.
 
Additional Manufacturer Narrative
The biosense webster, inc.Product analysis lab received the device for evaluation.The analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Biosense webster manufacturer's reference number (b)(4) has two reports: (1) mfr # 2029046-2022-01602 for product code bni35bdh (thermocool® sf nav bi-directional catheter) (2) mfr # 2029046-2022-01603 for product code bni35bdh (thermocool® sf nav bi-directional catheter).
 
Additional Manufacturer Narrative
Additional information was received on 01-aug-2022.The first ablation catheter¿s lot number was 30703467l, and the second was 30695591l.Both catheters had high impedance issue.No issue occurred on the smartablate generator.Transseptal puncture was conducted with rf needle.Based on this additional information, the concomitant product section was updated.Correct catheter settings were selected on the generator.No issue observed with pump.The pump was switching from ¿low¿ to ¿high¿ flow during ablation.No error message observed during the procedure.Df, realtime graph, vector force visualization features were used.Additional information was received on 8-aug-2022.It was reported that there was no char /coagulum/thrombus/clot was confirmed.The patient was anticoagulated, but act practice was unknown.The patient has not exhibited any neurological symptoms since the procedure was completed.Heparinized normal saline was used as the irrigation fluid.Visitag module was used with the respiration filter.The device evaluation was completed on 18-aug-2022.The product was returned to biosense webster (bwi) for evaluation.A visual inspection and an evaluation of all features of the device were performed following bwi procedures.Visual analysis revealed no damage or anomalies on the device.Per the event, several tests were performed.The magnetic, electrical, temperature, and force features were tested, and no issues were observed.In addition, the product was deflecting correctly.The device was partially irrigated; the device was dissected, and the distal holes of the dome were found obstructed with reddish material inside.The issue with the distal holes of the dome being occluded is considered not mdr reportable.The most likely consequence is an intraprocedural delay.The potential risk that it could cause or contribute to a serious injury or death is remote.A manufacturing record evaluation was performed for lot 30703467l, and no internal actions related to the complaint were found during the review.The root cause of the adverse event remains unknown; however, no device malfunction was found that could have caused this.The instruction for use (ifu) states that careful catheter manipulation must be performed to avoid cardiac damage, perforation, or tamponade.If the radiofrequency (rf) generator does not display temperature, verify that the appropriate cable is plugged into the rf generator.As part of biosense webster¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Biosense webster manufacturer's reference number pc-001143942 has two reports: (1) mfr # 2029046-2022-01602 for product code bni35bdh (thermocool® sf nav bi-directional catheter) (2) mfr # 2029046-2022-01603 for product code bni35bdh (thermocool® sf nav bi-directional catheter) on 01-aug-2022, additional information was received and the code of "char" (imdrf code of device contamination with body fluid (a180103)) has been removed since it was noted that the picture that was initially presented was from another issue not related to this complaint.Therefore, h6.Medical device problem code reflects the two codes that remain applicable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOCOOL SF CARTOXP,B-D,TC
Common Device Name
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
michelleann garcia
31 technology dr
irvine, CA 92618
646591-7981
MDR Report Key15009420
Report Number2029046-2022-01602
Device Sequence Number2395776
Product Code LPB
UDI-Device Identifier10846835003154
UDI-Public10846835003154
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P030031/S025
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/16/2025
Device Model NumberBNI35BDH
Device Catalogue NumberBNI35BDH
Device Lot Number30703467L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/06/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/15/2022
Supplement Date Received by Manufacturer08/01/2022
Initial Report FDA Received Date07/13/2022
Supplement Report FDA Received Date08/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/17/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SMARTABLATE GEN. KIT (JAPAN); SMARTABLATE GEN. KIT (JAPAN); THERMOCOOL SF CARTOXP,B-D,TC; UNK BRAND CABLE; UNK BRAND CABLE; UNK BRAND RADIOFREQUENCY (RF) NEEDLE; UNK JAPAN LIFELINE ABLAZE CATHETER
Outcome Attributed to Adverse Event Life Threatening; Required Intervention;
Patient SexMale
-
-