| Model Number |
MI1210 SYNCHRONY ST |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
They came to visit the clinic and reported that the user could not hear with the device.Re-implantation is considered.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: based on the received information from the field, a damage to the active electrode likely caused by device minute mobility is suspected.However, to determine an exact root cause device investigation of the explanted device is necessary.Re-implantation is considered but no date has been scheduled yet.
|
| |
|
Event or Problem Description
|
|
The user came to visit the clinic and reported that he could no longer hear with the device.Re-implantation is considered.
|
| |
|