Information was received from a healthcare provider (hcp) via a company representative regarding a patient receiving infumorph (10 mg/ml at 0.985 mg/day) via an implanted pump.The patient¿s medical history was failed back syndrome.The indication for pump use was spinal pain and non-malignant pain.It was reported that the patient had no pain relief in the last couple of months.There were no environmental, external, or patient factors that may have led or contributed to the issue.The hcp attempted a dye study but was unable to withdraw any fluid.The patient was taken to surgery.When the hcp made an incision at the distal part of the catheter, he noticed that the catheter had a complete cut.He thought that this was why the patient didn¿t have the expected relief.The catheter was replaced.The issue was noted to be resolved, and it was indicated that the hcp had no further information to provide regarding the event.
|
Concomitant medical products: product id 8731sc lot# serial# (b)(4), implanted: (b)(6) 2013, explanted: (b)(6) 2022 product type catheter.Other relevant device(s) are: product id: 8731sc, serial/lot #:(b)(4), ubd: (b)(6) 2015, udi#: (b)(4).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|