AIZU OLYMPUS CO., LTD. ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE Back to Search Results
Model Number LTF-S190-10
Device Problem Peeled/Delaminated (1454)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
The customer originally returned his olympus endoeye flex deflectable videoscope due to an unspecified air/water leakage.Additional details relating to the patient and the event have been requested, but no response has been received at this time.There was no patient harm or consequence reported as a result of this event.During the device evaluation it was discovered that adhesive on the object lens was peeling off.This report is being submitted to capture the peeling adhesive found during the device evaluation.
 
Manufacturer Narrative
The device was returned, and an initial evaluation was conducted by olympus; however, investigation is ongoing.During the initial evaluation, the user's report was confirmed, there were holes in the curved rubber section compromising watertightness.Additionally, as noted, the adhesive on the object lens was peeling off.There was noise noted in the image and scratches present on the grip, video connector, and light guide connector.If additional information becomes available following the device evaluation, a supplemental report will be filed.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the event was unable to be identified; however, the event most likely occurred due to physical stress due to dropping or bumping, or deterioration due to chemical stress.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE
Type of Device
FLEX DEFLECTABLE VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi fukushima
Manufacturer Contact
masaharu hirose
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8-520
JA   965-8520
426422891
MDR Report Key15013625
MDR Text Key296171861
Report Number9610595-2022-00003
Device Sequence Number1
Product Code FGB
UDI-Device Identifier04953170310386
UDI-Public04953170310386
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K111425
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 09/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/14/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberLTF-S190-10
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/22/2022
Was the Report Sent to FDA? No
Date Manufacturer Received09/02/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/21/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-