• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SWITZERLAND MANUFACTURING GMBH ANN, BLUNT TIP SCREW, 4X42MM; ROD, FIXATION, INTRAMEDULLARY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SWITZERLAND MANUFACTURING GMBH ANN, BLUNT TIP SCREW, 4X42MM; ROD, FIXATION, INTRAMEDULLARY Back to Search Results
Model Number 47-2486-042-40
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/16/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Concomitant medical products: cortical bone screw, 4x26mm; catalog#47248612640; lot#3091271.Cortical bone screw, 4x28mm; catalog#47248612840; lot#3082043.Proximal humerus nail cap, 0mm; catalog#47248801000; lot#3082050.Torque limiting handle; catalog#27923; lot#unknown.Report source: foreign: japan.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0009613350-2022-00369, 0009613350-2022-00371, 0009613350-2022-00372, 0009613350-2022-00373.
 
Event or Problem Description
It was reported that the surgeon found that the proximal screw was backed out from the proper position 1 month after the initial surgery.The surgeon will continue to monitor the patient's condition.Revision surgery is not currently planned.No further patient outcome was provided.Attempts have been made and no further information has been provided yet.
 
Additional Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated: b4, b5, g3, g6, h1, h2, h6, h10.As no product was returned, visual and dimensional evaluations could not be performed.A review of the device history records identified no deviations or anomalies during manufacturing.Devices are used for treatment.No surgical report or x-rays are available for investigation.Based on the investigation it could be assumed that further possible contributing factors to the migration of the screw might be multifactorial related to either patient condition, behavior or implantation procedure.If and to what extent any of these aspects may have influenced the migration of the screw remains unknown.The investigation also considered whether the design limitation of the corelock mechanism could be a potential contributing factor.However, further biomechanical testing identified that the product performance is in an acceptable range in comparison with legally marketed similar devices.In conclusion, as the cause may be multifactorial an exact root cause could not be identified for the migration of the screw.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANN, BLUNT TIP SCREW, 4X42MM
Common Device Name
ROD, FIXATION, INTRAMEDULLARY
Manufacturer (Section D)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ  8404
Manufacturer (Section G)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ   8404
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key15014034
Report Number0009613350-2022-00370
Device Sequence Number6855805
Product Code HSB
UDI-Device Identifier00889024505452
UDI-Public(01)00889024505452(17)260508(10)3073793
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K181827
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date05/08/2026
Device Model Number47-2486-042-40
Device Catalogue Number47-2486-042-40
Device Lot Number3073793
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/17/2022
Supplement Date Received by Manufacturer08/31/2022
Initial Report FDA Received Date07/14/2022
Supplement Report FDA Received Date09/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/07/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-