It was reported that on (b)(6) 2022, during post-op event bilateral revision of femur fixation done due to non-union and mal reduction.There was a patient consequence.On the right side there was the removal of four (4) screws from a plate, compression application and re-insertion of new screws into existing plate was done.On the left side there was removal of retrograde femoral nail with three (3) screws and new implantation of nail with compression plating was done.One screw was broken.This report is for one (1) rfna / 14 / 420 5 deg bend / sterile.This is report 1 of 4 for complaint.
|
|
Product complaint # (b)(4).Additional narrative: complainant part is not expected to be returned for manufacturer review/investigation.Investigation summary photo investigation: visual analysis of the photo revealed that there was no damage or defects with the rfna / 14 / 420 5 deg bend / sterile.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was not confirmed for rfna / 14 / 420 5 deg bend / sterile.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: manufacturing location: monument.Manufacturing date: 12-jan-2021.Expiration date: unknown.Part number: 04.233.442s, rfn/14mm/420mm 5 degree bend/pe inlay/sterile.Lot number: 76p7474 (sterile).Lot quantity: (b)(4).One piece was scrapped in cell at op #130, final inspection, for an unacceptable surface finish-dings/scratches.Production order traveler met all inspection acceptance criteria apart from the one piece noted.Inspection sheet, in-process / inspect dimensional / final ns498968 rev a met all inspection acceptance criteria.Inspection sheet, rfn assembly inspection, ns529720 rev b met all inspection acceptance criteria.Packaging bom was reviewed and determined to be conforming with no deviations to normal packaging identified.Note: there was no scn number recorded on the production order traveler.Therefore, the scn for this lot could not be reviewed.There was no pll included in this dhr.Therefore, the pll for this lot could not be reviewed and the expiration date could not be notated.This lot met all dimensional and visual criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.Component part(s) reviewed: component parts were not reviewed as the reported complaint condition of ¿removal due to non-union and mal reduction¿ does not indicate breakage of the nail.Therefore, review of the raw materials would not be pertinent to the reported complaint condition.This lot met all dimensional, visual and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|