| Model Number |
X-SPM-03 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A customer reported problems with the digital marker microscope, medical intervention was required: toric repositions in the unknown eye of a patient, during surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information provided in h.3., h.6., and h.10.A review of the device history record (dhr) traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.The root causes user - error.The verion supports a maximum tracking range of +-15° and for any reason if the system is pushed to work outside this range (levels of centration, zoom, illumination), this will lead to compromised performance of the verion system.Hence the flopping around of the axis.The manufacturer internal reference number is: (b)(4).
|
| |
|